Sputtering equipment for R&D QAM series-QAM-4C-S
Sputtering equipment for R&D QAM series-ULVAC Sales Co., Ltd.

Sputtering equipment for R&D QAM series
ULVAC Sales Co., Ltd.


About This Product

■Summary This is a device that can be used in a wide range of applications, based on ULVAC's extensive film-forming technology and experience. Achieved miniaturization while providing versatility and expandability. Process chambers and exhaust systems can be freely combined to suit your application. ■Applications Based on ULVAC's extensive film-forming technology and experience, this low-cost R&D equipment can handle a wide range of functions required for small-scale experimental applications such as all kinds of research and development and material development. ■Can be customized to suit your experiment/research purpose/budget We have built a unit that can be expanded by incorporating plenty of the technology and know-how we have cultivated over the years. For example, by adding each module to the base machine, it is possible to eventually create a 2-module machine (with L/L room). In addition, you can choose between mechanical and electrical integrated type and control module separated type to suit your layout. ■Achieve advanced functions, intuitive operation, and safety ・Safe and secure with intuitive operation ・Process data can be analyzed with peace of mind (data logging: optional) ・Recipe control enables film formation with good reproducibility ・Stable operation of equipment with interlock equivalent to production equipment ■Low pressure process It has a discharge sustaining pressure range of approximately 1Pa to 0.1Pa, allowing dense film formation by sputtering at a lower pressure than before. ■Multi-dimensional simultaneous/multi-layer sputter deposition Because each cathode aims at the center of the substrate, simultaneous multi-source sputtering is possible. Laminated films can be created using shutter control. Also, by using a jig (optional), static counter sputtering is also possible. ■Support for magnetic thin film The same cathode can be used with various magnetic materials (Fe, Ni, Co, etc.) ■Adoption of LTS By adopting LTS*1, it is possible to minimize the influence of non-uniform plasma near the cathode on the substrate. It is also possible to minimize the influence of the cathode magnet's magnetic field. Note) *1: LTS: Long Throw Sputter ■UHV compatible (QAM-4-S only) By using bakeable equipment, the ultimate pressure in the film forming chamber can be kept below 1 x 10-6Pa*2. Note) *2: Adopts a double O-ring seal structure (intermediate exhaust), greatly improving maintainability compared to conventional metal seals.

  • Product

    Sputtering equipment for R&D QAM series

Share this product


90+ people viewing

Last viewed: 10 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of Sputtering equipment for R&D QAM series

Product Image Part Number Price (excluding tax) Cathode method Chamber baking Compatible board size Compressed air Cooling water Earth Electricity Exhaust Exhaust system Film formation Film thickness distribution Maximum gas introduction amount Number of cathodes Onboard power supply Processing method Substrate heating mechanism Target size Ultimate pressure
Sputtering equipment for R&D QAM series-Part Number-QAM-4C-S

QAM-4C-S

Available upon quote Long throw magnetron sputtering
・Helicon sputter*1 *Optional
Bake at 110℃ (chamber water cooling included) *Optional Max.φ4 inch x 1 piece 0.5~0.7MPa 20~28℃, specific resistance 5KΩ or more, supply pressure 0.2~0.3MPa A class, D class 3φAC200V 50/60Hz Touch panel (PLC control) Remote manual*3 Fully automatic*4
*3: Manual/semi-automatic operation from touch panel
*4: Recipes such as power and time can be input using the touch panel.
1) Turbo pump
2) Rotary pump
Dry pump *Optional
Touch panel (PLC control) Remote manual*3 Fully automatic*4
*3: Manual/semi-automatic operation from touch panel
*4: Recipes such as power and time can be input using the touch panel.
Within ±5% (combined with substrate rotation during Al film formation) 1) Max.50sccm (Ar)
2) Max.10sccm (O2)
3) Max.10sccm (N2)
2 inch cathode (for both magnetic and non-magnetic materials)
Up to 4 units can be installed
DC500W 1 unit
・DC500W, RF300W*2 can be selected, and up to 2 units can be installed *Optional
Batch type Lamp heater Regular use 400℃ *Optional Φ2 inch 1×10-4Pa or less