Product
On-site PH/ORP meter FEB-102-PHHandling Company
Shinko Technos Co., Ltd.Categories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Ambient temperature | Calibration function | Case, panel | Conformity standard | Drip -proof / dustproof structure | EVT output (2 points) 1 | EVT output (2 points) 2 | EVT output (2 points) 3 | Entering 2 | External dimensions | Input 1 | Input sampling cycle | Linearness | Mounting method | Peripheral humidity | Repetition | Self -diagnostic function | Serial communication (Options: C5) | Serial communication (Options: C5) Communication method | Serial communication (option: C5) cable length | Serial communication (option: C5) communication line | Serial communication (option: C5) communication speed | Serial communication (option: C5) synchronous method | Temperature instruction accuracy | Time accuracy | Transmission output 1, 2-1 | Transmission output 1, 2-2 | mass | power consumption | power supply |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
FEB-102-pH, CS |
Available upon quote | -20 ~ 50 ° C (how to adapt to the instruction accuracy at 0 to 50 ° C and avoid direct sunlight) | ・ PH calibration 2 points Automatic proofreading or manual proofreading, type of standard solution PH2, 4, 7, 9, 10 (JIS) ・ Temperature school regular (1 point) |
・ Case: Polycarbonate color: Metallic gray ・ Panel: Membrane sheet |
ROHS command support | IP65 |
・ Setting accuracy: Same as instruction accuracy ・ Output operation: P operation proportional belt other than 0, when the ON/OFF operation proportional belt is set to 0. ・ Relay contact 1A |
・ Control capacity: 3A 250V AC (resistance load), 1A 250V AC (induced load COSφ = 0.4) ・ Electrical life: 100,000 times |
・ Operation ON delay timer: 0 to 10000 seconds ・ Operation OFF delay timer: 0 to 10000 seconds |
ORP composite electrode sensor (temperature element: CU500/25 ° C, PT100 or PT1,000) | W239.5xh190XD75mm | PH composite electrode sensor (PH sensor: JIS Z8802 compliant, temperature element: CU500/25 ° C, PT100 or PT1000) | 125 MS (2 input) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Wall installation method | 35 ~ 95%RH (but not condensation) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Motor the CPU with a watchdog timer and make the instrument into an initial state in the event of an abnormality. | Perform the following operations from the external computer.・ Reading and setting of various set values ・ Change, adjustment of functions / functions of ORP or status | Half double communication | 1.2 km (maximum) Cable resistance within 50Ω (end resistance: None or 120Ω or more on both sides) | EIA RS-485 compliant | Select 9600, 19200, 38400 BPS by key operation | Toning synchronization | ± 1 ℃ | Within ± 1%for the set time | PH, temperature, ORP value, and MV are converted to analog amount for each input sampling and output in current. (The decimal position is fixed regardless of the decimal position selection) | ・ Dusive ability: 12000 ・ Current: 4 ~ 20mA DC (maximum load resistance 550Ω) ・ Output accuracy: within ± 0.3%of transmission output span | Approximately 950g | Approximately 9 VA | 100 ~ 240 V AC 50/60 Hz tolerance fluctuation range: 85 ~ 264 v AC |
![]() |
FEB-102-PH, EVT3 |
Available upon quote | -20 ~ 50 ° C (how to adapt to the instruction accuracy at 0 to 50 ° C and avoid direct sunlight) |
・ PH calibration 2 points Automatic proofreading or manual proofreading, type of standard solution pH2, 4, 7, 9, 10 (JIS) ・ Temperature school regular (1 point) |
・ Case: Polycarbonate color: Metallic gray ・ Panel: Membrane sheet |
ROHS command support | IP65 |
・ Setting accuracy: Same as instruction accuracy ・ Output operation: P operation proportional belt other than 0, when the ON/OFF operation proportional belt is set to 0. ・ Relay contact 1A |
・ Control capacity: 3A 250V AC (resistance load), 1A 250V AC (induced load COSφ = 0.4) ・ Electrical life: 100,000 times |
・ Operation ON delay timer: 0 to 10000 seconds ・ Operation OFF delay timer: 0 to 10000 seconds |
ORP composite electrode sensor (temperature element: CU500/25 ° C, PT100 or PT1000) | W239.5xh190XD75mm | PH composite electrode sensor (PH sensor: JIS Z8802 compliant, temperature element: CU500/25 ° C, PT100 or PT1000) | 125 MS (2 input) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Wall installation method | 35 ~ 95%RH (but not condensation) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Motor the CPU with a watchdog timer and make the instrument into an initial state in the event of an abnormality. |
Perform the following operations from the external computer. ・ Read and set various set values ・ Reading pH, temperature, ORP value or state ・ Change and adjust function |
Half double communication | 1.2 km (maximum) Cable resistance within 50Ω (end resistance: None or 120Ω or more on both sides) | EIA RS-485 compliant | Select 9600, 19200, 38400 BPS by key operation | Toning synchronization | ± 1 ℃ | Within ± 1%for the set time | PH, temperature, ORP value, and MV are converted to analog amount for each input sampling and output in current. (The decimal position is fixed regardless of the decimal position selection) |
・ Resolution: 12,000 ・ Current: 4 ~ 20mA DC (maximum load resistance 550Ω) ・ Output accuracy: within ± 0.3%of transmission output span |
Approximately 950g | Approximately 9 VA | 100 ~ 240 V AC 50/60 Hz tolerance fluctuation range: 85 ~ 264 v AC |
![]() |
FEB-102-PH, EVT4 |
Available upon quote | -20 ~ 50 ° C (how to adapt to the instruction accuracy at 0 to 50 ° C and avoid direct sunlight) |
・ PH calibration 2 points Automatic proofreading or manual proofreading, type of standard solution pH2, 4, 7, 9, 10 (JIS) ・ Temperature school regular (1 point) |
・ Case: Polycarbonate color: Metallic gray ・ Panel: Membrane sheet |
ROHS command support | IP65 |
・ Setting accuracy: Same as instruction accuracy ・ Output operation: P operation proportional belt other than 0, when the ON/OFF operation proportional belt is set to 0. ・ Relay contact 1A |
・ Control capacity: 3A 250V AC (resistance load), 1A 250V AC (induced load COSφ = 0.4) ・ Electrical life: 100,000 times |
・ Operation ON delay timer: 0 to 10000 seconds ・ Operation OFF delay timer: 0 to 10000 seconds |
ORP composite electrode sensor (temperature element: CU500/25 ° C, PT100 or PT1000) | W239.5xh190XD75mm | PH composite electrode sensor (PH sensor: JIS Z8802 compliant, temperature element: CU500/25 ° C, PT100 or PT1000) | 125 MS (2 input) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Wall installation method | 35 ~ 95%RH (but not condensation) | PH meter: ± 0.05PH, ORP meter: within ± 5mv (in equivalent input) | Motor the CPU with a watchdog timer and make the instrument into an initial state in the event of an abnormality. |
Perform the following operations from the external computer. ・ Read and set various set values ・ Reading pH, temperature, ORP value or state ・ Change and adjust function |
Half double communication | 1.2 km (maximum) Cable resistance within 50Ω (end resistance: None or 120Ω or more on both sides) | EIA RS-485 compliant | Select 9600, 19200, 38400 BPS by key operation | Toning synchronization | ± 1 ℃ | Within ± 1%for the set time | PH, temperature, ORP value, and MV are converted to analog amount for each input sampling and output in current. (The decimal position is fixed regardless of the decimal position selection) | ・ Dusive ability: 12000 ・ Current: 4 ~ 20mA DC (maximum load resistance 550Ω) ・ Output accuracy: within ± 0.3%of transmission output span | Approximately 950g | Approximately 9 VA | 100 ~ 240 V AC 50/60 Hz tolerance fluctuation range: 85 ~ 264 v AC |
Click on the part number for more information about each product