Carbon sulfur analyzer CS744 type series CS744/S744Handling Company
Cosmo Trading Co., Ltd.Categories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis accuracy | Analysis scope | Analysis time | Calibration | Cycle time | Detection method | Dimensions (with touch screen monitor) | Furnace | Gas flow rate | Gas used | Operating environment | Option | Power capacity | Processing capacity | Reagent | Sample weight | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CS744 |
Available upon quote |
Carbon: Standard deviation (s) 0.001mg or RSD0.5% *Whichever is greater Sulfur: Standard deviation (s) 0.001mg or RSD1.5% *Whichever is greater |
Carbon: 0.002mg~60mg (2ppm~6%) Sulfur: 0.002mg~17.5mg (2ppm~1.75%) |
40 seconds (typical) | Standard sample (1 point or multiple points), automatic or manual | 130 seconds (typical) | Non-dispersive infrared absorption method | Width: 59 cm x Depth: 75 cm x Height: 84 cm | High frequency furnace 2.2kw |
Carrier: 3 L/min Drive: 1 L/min |
Carrier: Oxygen 99.5% (0.24 MPa) Drive: Compressed air or nitrogen (0.28 MPa; does not contain water or oil) |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
・Mounted touch screen monitor ・Auto cleaner and automatic combustion tube replacement system ・Oxygen gas purification pipe and heated flow controller |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 5,500BTU/hr | 27 samples/hour (typical) |
・Magnesium perchlorate ・Sodium hydroxide (coated on an inert substrate) ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1g | Body weight: 140kg (with touchscreen monitor), 132kg (without touchscreen monitor) | |
S744 |
Available upon quote |
Carbon: Standard deviation (s) 0.001mg or RSD0.5% *Whichever is greater Sulfur: Standard deviation (s) 0.001mg or RSD1.5% *Whichever is greater |
Carbon: 0.002mg~60mg (2ppm~6%) Sulfur: 0.002mg~17.5mg (2ppm~1.75%) |
40 seconds (typical) | Standard sample (1 point or multiple points), automatic or manual | 130 seconds (typical) | Non-dispersive infrared absorption method | Width: 59 cm x Depth: 75 cm x Height: 84 cm | High frequency furnace 2.2kw |
Carrier: 3 L/min Drive: 1 L/min |
Carrier: Oxygen 99.5% (0.24 MPa) Drive: Compressed air or nitrogen (0.28 MPa; does not contain water or oil) |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
・Mounted touch screen monitor ・Auto cleaner and automatic combustion tube replacement system ・Oxygen gas purification pipe and heated flow controller |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 5,500BTU/hr | 27 samples/hour (typical) |
・Magnesium perchlorate ・Sodium hydroxide (coated on an inert substrate) ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1g | Body weight: 140kg (with touchscreen monitor), 132kg (without touchscreen monitor) |