Product
Carbon/Sulfur Analyzer CS844 Series CS844/CS844ESHandling Company
Cosmo Trading Co., Ltd.Categories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis accuracy | Analysis scope | Analysis time | Calibration | Cycle time | Detection method | Dimensions (with touch screen monitor) | Furnace | Gas flow rate | Gas used | Internal cooling water | Operating environment | Operation | Power capacity | Processing capacity | Reagent | Sample weight | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CS844 |
Available upon quote |
Carbon: Standard deviation (s) 0.0003mg (0.3ppm) or RSD0.5% *Whichever is greater Sulfur: Standard deviation (s) 0.0003mg (0.3ppm) or RSD0.5% *Whichever is greater |
Carbon: 0.0006mg~60mg (0.6ppm~6.0% for 1g sample) Sulfur: 0.0006mg~60mg (0.6ppm~6.0% for 1g sample) |
40 seconds (typical) | Standard sample (1 point or multiple points), automatic or manual | 130 seconds (typical) | Non-dispersive infrared absorption method | Width: 59 cm x Depth: 75 cm x Height: 84 cm | High frequency furnace, maximum 2.2kw (heating 0~100% power), water cooling type |
Carrier: 3 L/min Drive: 1 L/min |
Carrier: Oxygen 99.5% (0.24 MPa) Drive: Compressed air or nitrogen (0.28 MPa; no moisture or oil) Gas dosing: Carbon dioxide 99.99% (0.14 MPa) |
300mL, LECO Coolant |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
- | Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 5,500BTU/hr | 27 samples/hour (typical) |
・Magnesium perchlorate ・Sodium hydroxide (coated on an inert substrate) ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1g | Body weight: 140kg (with touchscreen monitor), 132kg (without touchscreen monitor) | |
CS844ES |
Available upon quote |
Carbon: Standard deviation (s) 0.3ppm or RSD0.5% *Whichever is larger Sulfur (standard mode): Standard deviation (s) 0.3ppm or RSD0.5% *Whichever is larger |
Carbon: 0.0006mg~60mg (0.6ppm~6%) Sulfur (standard mode): 0.0006mg~60mg (0.6ppm~6%) |
Standard mode: 40 seconds, ES mode: 130 seconds | Standards (single or multipoint), automatic or manual | Standard mode: 130 seconds, ES mode: 270 seconds | Non-dispersive infrared absorption method | Height: 84 cm, Width: 64 cm, Depth: 75 cm (including mounted touch screen) | High frequency furnace, maximum 2.2kw (0~100% output control), water cooling type |
Carrier: 3L/min Drive: 1L/min |
Carrier gas: Oxygen (99.5%), 0.24 MPa ±10% Driving gas: Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10%, etc. |
300mL, LECO Coolant |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
Windows software | Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 5,500BTU/hr | - |
・Magnesium perchlorate ・Sodium hydroxide (coated on an inert substrate) ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1.0g | Body 140kg (including monitor), 132kg (without monitor) |
Click on the part number for more information about each product