Product
Organic analysis sulfur carbon analyzer SC832 typeHandling Company
LECO Japan LLCCategories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis accuracy *Whichever is greater | Analysis range (sample weight 200 mg) | Analysis time | Detection method | Furnace | Gas used | Heater section | Option | Part number 1 | Part number 2 | Power capacity | Reagents used | Sample weight | Size | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
SC832 |
Available upon quote |
Sulfur: Standard deviation (s) 0.004 mg or RSD 1% Sulfur (DR): Standard deviation (s) 0.0018 mg or RSD 1% Carbon: Standard deviation (s) 0.0088 mg or RSD 1% |
Sulfur: 0.008 mg~30 mg (0.004%~15%) Carbon: 0.0175 mg~200 mg (0.009%~100%) |
60~120 seconds (typical) | Non-dispersive infrared absorption method |
832/832DR: Resistance heating furnace 600℃~1450℃ (set value ±10℃) 832HT: Resistance heating furnace 600℃~1550℃ (set value ±10℃) |
Oxygen 99.5%, 0.1 MPa ±10%, 3.5 L/min, 1.0 L/min/purge | Made of molybdenum disilicide / 4 pieces |
Performance package (P); Leak check function and back pressure adjustment function Desktop type monitors are also available. |
S832MC: Sulfur analyzer C832MC: Carbon analyzer SC832MC: Sulfur carbon analyzer |
S832DR-MC: Dual range sulfur analyzer SC832DR-MC: Dual range sulfur carbon analyzer |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 14 A (maximum), 5 A (standby), 5,100 BTU/hr (during analysis), 3,100 BTU/hr (standby) | Magnesium perchlorate (Anhydron) | Normal Sulfur only: 350 mg (coal), Carbon/Sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm (66 cm) x Height: 80 cm | 88 kg (with mount monitor), 78 kg (without mount monitor) |
![]() |
SC832DR |
Available upon quote |
Sulfur: Standard deviation (s) 0.004 mg or RSD 1% Sulfur (DR): Standard deviation (s) 0.0018 mg or RSD 1% Carbon: Standard deviation (s) 0.0088 mg or RSD 1% |
Sulfur: 0.0035 mg~98 mg (0.002%~49%) Carbon: 0.0175 mg**~200 mg (0.009%~100%) |
60~120 seconds (typical) | Non-dispersive infrared absorption method |
832/832DR: Resistance heating furnace 600℃~1450℃ (set value ±10℃) 832HT: Resistance heating furnace 600℃~1550℃ (set value ±10℃) |
Oxygen 99.5%, 0.1 MPa ±10%, 3.5 L/min, 1.0 L/min/purge | Made of molybdenum disilicide / 4 pieces |
Performance package (P); Leak check function and back pressure adjustment function Desktop type monitors are also available. |
S832MC: Sulfur analyzer C832MC: Carbon analyzer SC832MC: Sulfur carbon analyzer |
S832DR-MC: Dual range sulfur analyzer SC832DR-MC: Dual range sulfur carbon analyzer |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 14 A (maximum), 5 A (standby), 5,100 BTU/hr (during analysis), 3,100 BTU/hr (standby) | Magnesium perchlorate (Anhydron) | Normal Sulfur only: 350 mg (coal), Carbon/Sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm (66 cm) x Height: 80 cm | 88 kg (with mount monitor), 78 kg (without mount monitor) |
![]() |
S832 |
Available upon quote |
Sulfur: Standard deviation (s) 0.004 mg or RSD 1% Sulfur (DR): Standard deviation (s) 0.0018 mg or RSD 1% Carbon: Standard deviation (s) 0.0088 mg or RSD 1% |
- | 60~120 seconds (typical) | Non-dispersive infrared absorption method |
832/832DR: Resistance heating furnace 600℃~1450℃ (set value ±10℃) 832HT: Resistance heating furnace 600℃~1550℃ (set value ±10℃) |
Oxygen 99.5%, 0.1 MPa ±10%, 3.5 L/min, 1.0 L/min/purge | Made of molybdenum disilicide / 4 pieces |
Performance package (P); Leak check function and back pressure adjustment function Desktop type monitors are also available. |
S832MC: Sulfur analyzer C832MC: Carbon analyzer SC832MC: Sulfur carbon analyzer |
S832DR-MC: Dual range sulfur analyzer SC832DR-MC: Dual range sulfur carbon analyzer |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 14 A (maximum), 5 A (standby), 5,100 BTU/hr (during analysis), 3,100 BTU/hr (standby) | Magnesium perchlorate (Anhydron) | Normal Sulfur only: 350 mg (coal), Carbon/Sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm (66 cm) x Height: 80 cm | 88 kg (with mount monitor), 78 kg (without mount monitor) |
![]() |
S832DR |
Available upon quote |
Sulfur: Standard deviation (s) 0.004 mg or RSD 1% Sulfur (DR): Standard deviation (s) 0.0018 mg or RSD 1% Carbon: Standard deviation (s) 0.0088 mg or RSD 1% |
Sulfur (DR): 0.0035 mg~98 mg (0.002%~49%) Carbon: 0.0175 mg~200 mg (0.009%~100%) |
60~120 seconds (typical) | Non-dispersive infrared absorption method |
832/832DR: Resistance heating furnace 600℃~1450℃ (set value ±10℃) 832HT: Resistance heating furnace 600℃~1550℃ (set value ±10℃) |
Oxygen 99.5%, 0.1 MPa ±10%, 3.5 L/min, 1.0 L/min/purge | Made of molybdenum disilicide / 4 pieces |
Performance package (P); Leak check function and back pressure adjustment function Desktop type monitors are also available. |
S832MC: Sulfur analyzer C832MC: Carbon analyzer SC832MC: Sulfur carbon analyzer |
S832DR-MC: Dual range sulfur analyzer SC832DR-MC: Dual range sulfur carbon analyzer |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 14 A (maximum), 5 A (standby), 5,100 BTU/hr (during analysis), 3,100 BTU/hr (standby) | Magnesium perchlorate (Anhydron) | Normal Sulfur only: 350 mg (coal), Carbon/Sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm (66 cm) x Height: 80 cm | 88 kg (with mount monitor), 78 kg (without mount monitor) |
![]() |
C832 |
Available upon quote |
Sulfur: Standard deviation (s) 0.004 mg or RSD 1% Sulfur (DR): Standard deviation (s) 0.0018 mg or RSD 1% Carbon: Standard deviation (s) 0.0088 mg or RSD 1% |
Carbon: 0.0175 mg~200 mg (0.009%~100%) | 60~120 seconds (typical) | Non-dispersive infrared absorption method |
832/832DR: Resistance heating furnace 600℃~1450℃ (set value ±10℃) 832HT: Resistance heating furnace 600℃~1550℃ (set value ±10℃) |
Oxygen 99.5%, 0.1 MPa ±10%, 3.5 L/min, 1.0 L/min/purge | Made of molybdenum disilicide / 4 pieces |
Performance package (P); Leak check function and back pressure adjustment function Desktop type monitors are also available. |
S832MC: Sulfur analyzer C832MC: Carbon analyzer SC832MC: Sulfur carbon analyzer |
S832DR-MC: Dual range sulfur analyzer SC832DR-MC: Dual range sulfur carbon analyzer |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 14 A (maximum), 5 A (standby), 5,100 BTU/hr (during analysis), 3,100 BTU/hr (standby) | Magnesium perchlorate (Anhydron) | Normal Sulfur only: 350 mg (coal), Carbon/Sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm (66 cm) x Height: 80 cm | 88 kg (with mount monitor), 78 kg (without mount monitor) |
Click on the part number for more information about each product