Organic elemental analyzer CNS928 type-CNS928
Organic elemental analyzer CNS928 type-LECO Japan LLC

Organic elemental analyzer CNS928 type
LECO Japan LLC

LECO Japan LLC's Response Status

Response Rate

100.0%

Response Time

126.8hours


About This Product

■Carbon/nitrogen/sulfur analyzer using combustion method that can collect up to 3g of sample

Equipped with the latest hardware and a mounted touchscreen monitor, the 928 Series is easy to operate using combustion (modified Dumas) techniques for the widest range of applications including food, feed, fertilizer and soil. This is a device for measuring. The basic analysis principle and performance follow the conventional LECO macro combustion method device, and further improvements have been made in three areas: throughput, downtime reduction, and reliability. It is designed for maximum performance, with outstanding features such as combustion capacity for up to 3 g of sample (FP928 model), fast analysis times, and low running costs.

■Analysis target

Organic matter such as food, feed, soil, fertilizer, etc. Features and benefits

■High-throughput analysis and minimal downtime

・FP928 achieves cycle time of 5.3 minutes ・Extends reagent life and enables 4,000 continuous analyzes of reduced copper (3cc aliquot loop) ・Equipped with 100 position autoloader as standard

■Versatile and durable design suitable for a wide range of applications

・Horizontal ceramic combustion tube with a maximum temperature of 1,450°C completely burns macro samples in a pure oxygen stream ・A reusable sample boat with a wide opening facilitates sampling and supports efficient combustion. ・The ash content of the sample is collected together with the boat, reducing the frequency of combustion tube maintenance. ・Carrier gas can be compatible with both helium and argon, and can be switched without replacing hardware.

■Low running cost

・Highly efficient combustion furnaces do not need to be filled with reagents such as oxidizers, and intelligent control allows for longer lifespans of combustion tubes and heating elements. ・Remove water without using reagents using thermoelectric cooler (FP/CN model) ・Since a fixed amount of measurement gas is sampled from the ballast tank (aliquot system), consumption of reagents such as reduced copper is constant and minimal, regardless of sample volume, components, carbon content, etc. ・Energy-saving design reduces power supply capacity to less than half that of conventional models

  • Product

    Organic elemental analyzer CNS928 type

Share this product


160+ people viewing

Last viewed: 3 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

3 Models of Organic elemental analyzer CNS928 type

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Power capacity Weight Sample amount Furnace Detection method Dimensions body and autoloader Operating environment Analysis accuracy Carrier gas 《3 cm3 loop》 Analysis accuracy Carrier gas《10 cm3 loop》 Analysis range Carrier gas 《3 cm3 loop》 Analysis range Carrier gas《10 cm3 loop》 Gas used Gas used Combustion Gas carrier used Cycle time (EDTA, normal sample amount) Autoloader
Organic elemental analyzer CNS928 type-Part Number-CNS928

CNS928

Available upon quote

200 V (+10%/-15%, boosted to 230 V), 50/60 Hz, single phase 14 A (maximum), 6 A (standby), 4,100 BTU/hr during analysis

Main unit, autoloader approximately 178 kg Monitor 10 kg

Nitrogen (FP) up to 3.0 g, typical 1.0 g
Carbon/Nitrogen up to 1.0 g, typically 0.5 g
Carbon/Nitrogen/Sulfur max. 0.3 g, typically 0.2 g

Resistance heating furnace 600℃ – 1,450℃ ±10℃

Nitrogen: thermal conductivity method
Carbon/Sulfur: Non-dispersive infrared absorption method

Height: 112 cm x Width: 150 cm x Depth: 76 cm
(The length from the back panel of the device to the front leg is 56 cm)

Temperature: 15℃~35℃
Relative humidity: 20%~80% (non-condensing)

Helium Sulfur 0.02 mg or 0.8% RSD
Argon Sulfur 0.02 mg or 0.8% RSD

Helium Sulfur 0.01 mg or 0.4% RSD
Argon Sulfur 0.01 mg or 0.4% RSD

Helium Sulfur 0.04 mg~50 mg
Argon Sulfur 0.04 mg~50 mg

Helium Sulfur 0.02 mg~50 mg
Argon Sulfur 0.02 mg~50 mg

Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10%

Oxygen (99.9%), 0.17 MPa ±10%

Helium (99.99%), 0.17 MPa ±10%
Argon (99.99%), 0.17 MPa ±10%

5 minutes (12 samples/hour)

100 samples

Organic elemental analyzer CNS928 type-Part Number-FP928

FP928

Available upon quote

200 V (+10%/-15%, boosted to 230 V), 50/60 Hz, single phase 14 A (maximum), 6 A (standby), 4,100 BTU/hr during analysis

Main unit, autoloader approximately 178 kg Monitor 10 kg

Nitrogen (FP) up to 3.0 g, typical 1.0 g
Carbon/Nitrogen up to 1.0 g, typically 0.5 g
Carbon/Nitrogen/Sulfur max. 0.3 g, typically 0.2 g

Resistance heating furnace 600℃ – 1,450℃ ±10℃

Nitrogen: thermal conductivity method
Carbon/Sulfur: Non-dispersive infrared absorption method

Height: 112 cm x Width: 150 cm x Depth: 76 cm
(The length from the back panel of the device to the front leg is 56 cm)

Temperature: 15℃~35℃
Relative humidity: 20%~80% (non-condensing)

Helium Nitrogen 0.02 mg or 0.6% RSD
Argon Nitrogen 0.06 mg or 1.2% RSD

Helium Nitrogen 0.01 mg or 0.3% RSD
Argon Nitrogen 0.03 mg or 0.6% RSD

Helium Nitrogen 0.04 mg~300 mg
Argon Nitrogen 0.12 mg~300 mg

Helium Nitrogen 0.02 mg~300 mg
Argon Nitrogen 0.06 mg~300 mg

Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10%

Oxygen (99.9%), 0.17 MPa ±10%

Helium (99.99%), 0.17 MPa ±10%
Argon (99.99%), 0.17 MPa ±10%

5 minutes (12 samples/hour)

100 samples

Organic elemental analyzer CNS928 type-Part Number-CN928

CN928

Available upon quote

200 V (+10%/-15%, boosted to 230 V), 50/60 Hz, single phase 14 A (maximum), 6 A (standby), 4,100 BTU/hr during analysis

Main unit, autoloader approximately 178 kg Monitor 10 kg

Nitrogen (FP) up to 3.0 g, typical 1.0 g
Carbon/Nitrogen up to 1.0 g, typically 0.5 g
Carbon/Nitrogen/Sulfur max. 0.3 g, typically 0.2 g

Resistance heating furnace 600℃ – 1,450℃ ±10℃

Nitrogen: thermal conductivity method
Carbon/Sulfur: Non-dispersive infrared absorption method

Height: 112 cm x Width: 150 cm x Depth: 76 cm
(The length from the back panel of the device to the front leg is 56 cm)

Temperature: 15℃~35℃
Relative humidity: 20%~80% (non-condensing)

Helium Carbon 0.02 mg or 0.8% RSD
Argon Carbon 0.02 mg or 0.8% RSD

Helium Carbon 0.01 mg or 0.4% RSD
Argon Carbon 0.01 mg or 0.4% RSD

Helium Carbon 0.04 mg~200 mg
Argon Carbon 0.04 mg~200 mg

Helium Carbon 0.02 mg~200 mg
Argon Carbon 0.02 mg~200 mg

Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10%

Oxygen (99.9%), 0.17 MPa ±10%

Helium (99.99%), 0.17 MPa ±10%
Argon (99.99%), 0.17 MPa ±10%

5 minutes (12 samples/hour)

100 samples

Click on the part number for more information about each product

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Elemental Analyzers Products

Other products of LECO Japan LLC

Reviews shown here are reviews of companies.


View more products of LECO Japan LLC

About Company Handling This Product

Response Rate

100.0%


Response Time

126.8hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree