Product
Inorganic analysis oxygen/nitrogen/hydrogen analyzer ONH836 typeHandling Company
LECO Japan LLCCategories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis range (sample weight 1g) | Analysis time 1 | Analysis time 2 | Analytical accuracy (sample weight 1g) 1 | Analytical accuracy (sample weight 1g) 2 | Calibration | Detection method Nitrogen | Detection method Oxygen, hydrogen | Dimensions Depth | Dimensions Width | Dimensions height | Furnace | Gas carrier used | Gas flow rate | Gas options | Gas used | Internal cooling water | Operating environment temperature | Part number 1 | Part number 2 | Power capacity | Reagent | Sample weight | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ONH836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
ON836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
OH836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
NH836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
O836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
N836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) | |
H836 |
Available upon quote |
Oxygen: 0.05 ppm~5.0% Nitrogen (carrier gas helium): 0.05 ppm~3.0% Nitrogen (carrier gas argon): 0.2 ppm~3.0% Hydrogen: 0.1 ppm~2500 ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds Nitrogen [carrier gas helium]: 100 seconds, [carrier gas argon]: 130 seconds |
Hydrogen [carrier gas helium]: 90 seconds, [carrier gas argon]: 100 seconds Cycle time [carrier gas helium]: 180 seconds Cycle time [carrier gas argon]: 210 seconds |
Oxygen: Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025 ppm or RSD 0.3% *Whichever is larger |
Nitrogen (carrier gas argon): Standard deviation (s) 0.1 ppm or RSD 0.3% *Whichever is larger Hydrogen: Standard deviation (s) 0.05 ppm or RSD 2% *Whichever is larger |
Standard material (1 point or multiple points); Manual; Gas dose | Thermal conductivity method | Non-dispersive infrared absorption method |
With touch screen monitor (85 cm) No touchscreen monitor (76 cm) |
71 cm | Load head cover with lift (100 cm), normal (91.5 cm) 71 cm | Electrode furnace, current/power control up to 7500 watts, water-cooled | Helium (99.99%), 0.15 MPa or Argon (99.999%), 0.15 MPa |
Carrier: 490 cc/min Drive: 280 cc/analysis |
Gas dose: carbon dioxide (99.99%) 0.14 MPa, nitrogen (99.99%) 0.14 MPa | Dry nitrogen, 0.28 MPa (no water or oil) | Distilled water 3.2 L |
Temperature: 15~35℃ Relative humidity: 20~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, mount type monitor |
[ONH836-C] Oxygen/nitrogen/hydrogen analyzer, desktop type monitor [ONH836-HC] Oxygen/nitrogen/hydrogen analyzer, auto cleaner, desktop type monitor |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 60 A, 12,500 BTU/hr *Average when operating under normal settings. |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal reagent (OMI tube) |
Usually 1 g | Body: 186 kg (without touch screen monitor) |