Product
Inorganic analysis carbon sulfur analyzer CS844 typeHandling Company
LECO Japan LLCCategories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis accuracy carbon | Analysis accuracy sulfur | Analysis range Sulfur | Analysis range carbon | Analysis time | Calibration | Cycle time | Detection method | Dimensions (with touch screen monitor) | Furnace | Gas carrier used | Gas flow rate | Gas used | Gas used Gas dosing | Internal cooling water | Operating environment | Operation | Power capacity | Processing capacity | Reagent | Sample weight | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CS844 |
Available upon quote | Standard deviation (s) 0.0003 mg (0.3 ppm) or RSD 0.5% *Whichever is greater | Standard deviation (s) 0.0003 mg (0.3 ppm) or RSD 0.5% *Whichever is greater | 0.0006 mg~60 mg (0.6 ppm~6.0% for 1 g sample) | 0.0006 mg~60 mg (0.6 ppm~6.0% for 1 g sample) | 40 seconds (typical) | Standard sample (1 point or multiple points), automatic or manual | 130 seconds (typical) | Non-dispersive infrared absorption method | Width: 59 cm x Depth: 75 cm x Height: 84 cm | High frequency furnace, maximum 2.2 kw (heating 0~100% power), water cooling type | Oxygen 99.5% (0.24 MPa) |
Carrier: 3 L/min Drive: 1 L/min |
Compressed air or nitrogen (0.28 MPa; no moisture or oil) | Carbon dioxide 99.99% (0.14 MPa) | 300mL, LECO Coolant |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
- | Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 30 A, 5,500 BTU/hr | 27 samples/hour (typical) |
・Magnesium perchlorate ・Sodium hydroxide (coding on inert substrate) ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1 g | Body weight 140 kg (with touch screen monitor), 132 kg (without touch screen monitor) | |
CS844ES |
Available upon quote | Standard deviation (s) 0.3 ppm or RSD 0.5% *Whichever is larger |
(Standard mode): Standard deviation (s) 0.3 ppm or RSD 0.5% *Whichever is larger (ES mode): Standard deviation (s) 0.05 ppm or RSD 0.5% *Whichever is larger |
(Standard mode): 0.0006 mg~60 mg (0.6 ppm~6%) (ES mode): 0.0001 mg~0.05 mg (0.1 ppm~50 ppm) |
0.0006 mg~60 mg (0.6 ppm~6%) | Standard mode: 40 seconds ES mode: 130 seconds | Standards (single or multipoint), automatic or manual | Standard mode: 130 seconds ES mode: 270 seconds | Non-dispersive infrared absorption method | Height: 84 cm, Width: 64 cm, Depth: 75 cm (including mounted touchscreen) | High frequency furnace, maximum 2.2 kW (0-100% output control), water-cooled | Oxygen (99.5%), 0.24 MPa ±10% |
Carrier gas: 3 L/min (ES mode: 0.8 L/min) Driving gas: 1 L/min |
Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10% | Carbon dioxide (99.99%), 0.14 MPa ±10% | 300mL, LECO Coolant |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
Windows software | Single phase 200 V~ (+10/-15%, boosted to 230 V), 50/60 Hz, 30 A, 5,500 BTU/hr | - |
・Sodium hydroxide (coated on an inert substrate) ・Magnesium perchlorate ・Platinum silica gel ・Rare earth based copper oxide ·cellulose |
Usually 1.0 g | Body: 140 kg (including monitor), 132 kg (without monitor) |