All Categories
History
Response Rate
100.0%
Response Time
126.8hours
Product
Inorganic analysis carbon sulfur analyzer CS844 typeHandling Company
LECO Japan LLCCategories
Click on the part number for more information about each product
Image | Part Number | Price (excluding tax) | Power capacity | Weight | Reagent | Sample weight | Furnace | Detection method | Dimensions (with touch screen monitor) | Operating environment | Analysis accuracy sulfur | Analysis accuracy carbon | Analysis range Sulfur | Analysis range carbon | Analysis time | Internal cooling water | Gas used | Gas carrier used | Gas used Gas dosing | Cycle time | Calibration | Gas flow rate | Operation | Processing capacity |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CS844 |
Available upon quote |
Single phase 200 V~ (+10/-15%: maximum output, boosted to 230 V), 50/60 Hz, 30 A, 5,500 BTU/hr |
Body weight 140 kg (with touch screen monitor), 132 kg (without touch screen monitor) |
・Magnesium perchlorate |
Usually 1 g |
High frequency furnace, maximum 2.2 kw (heating 0~100% power), water cooling type |
Non-dispersive infrared absorption method |
Width: 59 cm x Depth: 75 cm x Height: 84 cm |
Temperature: 15℃~35℃ |
Standard deviation (s) 0.0003 mg (0.3 ppm) or RSD 0.5% *Whichever is greater |
Standard deviation (s) 0.0003 mg (0.3 ppm) or RSD 0.5% *Whichever is greater |
0.0006 mg~60 mg (0.6 ppm~6.0% for 1 g sample) |
0.0006 mg~60 mg (0.6 ppm~6.0% for 1 g sample) |
40 seconds (typical) |
300mL, LECO Coolant |
Compressed air or nitrogen (0.28 MPa; no moisture or oil) |
Oxygen 99.5% (0.24 MPa) |
Carbon dioxide 99.99% (0.14 MPa) |
130 seconds (typical) |
Standard sample (1 point or multiple points), automatic or manual |
Carrier: 3 L/min |
- |
27 samples/hour (typical) |
|
CS844ES |
Available upon quote |
Single phase 200 V~ (+10/-15%, boosted to 230 V), 50/60 Hz, 30 A, 5,500 BTU/hr |
Body: 140 kg (including monitor), 132 kg (without monitor) |
・Sodium hydroxide (coated on an inert substrate) |
Usually 1.0 g |
High frequency furnace, maximum 2.2 kW (0-100% output control), water-cooled |
Non-dispersive infrared absorption method |
Height: 84 cm, Width: 64 cm, Depth: 75 cm (including mounted touchscreen) |
Temperature: 15℃~35℃ |
(Standard mode): Standard deviation (s) 0.3 ppm or RSD 0.5% *Whichever is larger |
Standard deviation (s) 0.3 ppm or RSD 0.5% *Whichever is larger |
(Standard mode): 0.0006 mg~60 mg (0.6 ppm~6%) |
0.0006 mg~60 mg (0.6 ppm~6%) |
Standard mode: 40 seconds ES mode: 130 seconds |
300mL, LECO Coolant |
Compressed air or nitrogen (no moisture or oil), 0.28 MPa ±10% |
Oxygen (99.5%), 0.24 MPa ±10% |
Carbon dioxide (99.99%), 0.14 MPa ±10% |
Standard mode: 130 seconds ES mode: 270 seconds |
Standards (single or multipoint), automatic or manual |
Carrier gas: 3 L/min (ES mode: 0.8 L/min) |
Windows software |
- |
Click on the part number for more information about each product
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
126.8hrs