Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha-JMS-T2000GC
Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha-Meishin Koki Co., Ltd.

Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha
Meishin Koki Co., Ltd.


About This Product

■Summary Two key technologies realized with the concept of high performance, simplicity and speed. Features ■New generation automatic analysis software Anyone can easily perform qualitative analysis of components not registered in the library. ・Automatic qualitative analysis combining EI method + various soft ionization method data ・Structural analysis using two AIs ・Deconvolution function ・Group analysis function ・Two-sample comparison (difference analysis) function ・Can be analyzed using EI data alone ■Newly developed hardware Achieves high basic performance that can support trace qualitative analysis and quantitative analysis of mixed samples. We will remove the limitations of analysis that you have given up on until now. ・Wide dynamic range: 4 digits *Measures OFN (octafluoronaphthalene) ・Device detection limit: IDL=18.7fg *Measuring OFN (octafluoronaphthalene) ・High-speed data acquisition: 50Hz ・High mass resolution: 30,000 ・High mass accuracy: 1ppm* Measuring methyl stearate ・Wide mass range: ~m/z 6,000 ■Newly developed hardware pursuing high performance AccuTOF™ GC-Alpha is the 6th generation GC-TOFMS from the original AccuTOF™ GC. The ion optical system is now in the opposite shape from the traditional V-shape. The name "Alpha" in the device name refers to the ion trajectory (inverted V-shaped = A) when viewed from the side of the device, and the developer's desire to start a new world of mass spectrometry with this device. It is being The AccuTOF™ GC-Alpha is an orthogonally accelerated time-of-flight mass spectrometer (OA-TOFMS) equipped with a two-stage field gradient reflectron. It has an ideal ion optical system that combines high ion transmittance (=sensitivity) and ultra-high resolution. ■New generation automatic analysis software that pursues simplicity and speed We have devised a new workflow for identifying "unknown compounds" using GC-MS. In this workflow, analysis is performed by integrating data obtained by the EI (electron ionization) method, which is a hard ionization method, and data obtained by soft ionization methods, such as the FI method. The automatic analysis software "msFineAnalysis" equipped with this new workflow was launched in 2018 as a software with a completely new concept, and has been used by many customers.

  • Product

    Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha

Share this product


60+ people viewing


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha

Product Image Part Number Price (excluding tax) Analysis department Data collection system Detection part Environmental magnetic field Exhaust equipment Floor vibration Gas Gas matograph Humidity Ion source Main options Maximum calorific value Power supply Room temperature Temperature change
Chromatograph High performance gas chromatograph time-of-flight mass spectrometer JMS-T2000GC AccuTOF™ GC-Alpha-Part Number-JMS-T2000GC

JMS-T2000GC

Available upon quote Reflectron time-of-flight mass spectrometer Continuous averager Dual microchannel plate AC magnetic field 1×10^-6T or less
Fixed magnetic field 5×10^-4T or less
Rotary pump (RP) exhaust equipment is required. Amplitude (p-p) 25μm or less
Acceleration 0.1m/s2 or less
Nitrogen gas: 0.5~0.6MPa, purity 97% or more
Helium gas: 0.5~0.68MPa, purity 99.999% or more. Hydrocarbon content less than 0.5ppm
Agilent8890 30~70% (no condensation) Electron ionization EI/FI/FD shared ion source, EI/PI shared ion source, CI ion source, FD/FI ion source, DIP, DEP 25,776 kJ/h (value calculated from the maximum power consumption of the main unit, gas chromatograph, and data system) Main unit: Single phase 200V±10%, 20A
Gas chromatograph: Single phase 200V±10%, 20A
Data system: Single phase 100V±10%, 15A
20~27℃ Within ±3℃/h