Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P-FIA-300P
Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P-Nittoseiko Analytech Co., Ltd

Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P
Nittoseiko Analytech Co., Ltd


About This Product

■Measurement items Phosphate phosphorus, total phosphorus FIA-300P is a device that uses the FIA-300 series' Flow Injection Analysis (FIA) method. Measurements can be made using the FIA ​​method as stipulated by JIS K0126 General Rules for Flow Analysis. The FIA ​​method is an analysis method in which a sample or reagent is introduced into the flow of the reagent or sample in a capillary tube, reaction operations are performed, and then the analytical components are detected and quantified using a detection section installed downstream. . Measurements are performed using a closed system that is less susceptible to external factors, allowing for highly accurate measurements. Additionally, since it prevents chemical substances from scattering, it is an environment-friendly and analyzer-friendly device. This device uses FIA's method to automate the total phosphorus analysis method stipulated in the 5th Total Volume Regulation, environmental standards, wastewater standards, tap water quality standards, and sewage discharge standards. Features ■High sensitivity analysis ・Pump with excellent liquid delivery stability ・Piping of constant length and constant temperature ・Since a stable baseline is obtained, small changes can be detected, and sufficient analytical sensitivity can be obtained with a small amount of sample. ■Piping configuration The simple piping configuration makes it easy to check the condition of the piping and to replace it easily. ■Small sample/small reagent Compared to manual analysis, the amount of samples and reagents used for analysis can be significantly reduced. As a result, the amount of waste fluid generated after analysis can also be reduced. ■Automation By connecting the auto sample changer ASC-F50L, you can automate the measurement of 50 to 150 samples. By selecting options, it is possible to handle samples with a large amount of suspended solids (SS) or samples with high concentrations.

  • Product

    Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P

Share this product


40+ people viewing


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P

Product Image Part Number Price (excluding tax) Analysis items Analysis method Corresponding standards External dimensions Heating section Liquid sending part Mass Measurement wavelength
Flow injection analyzer (FIA) Total phosphorus analyzer FIA-300P-Part Number-FIA-300P

FIA-300P

Available upon quote Phosphate phosphorus
Total phosphorus
Oxidative thermal decomposition - spectrophotometry (total phosphorus)
Absorption photometry (phosphate phosphorus)
JIS K 0102 46.1.1 Molybdenum blue spectrophotometry
JIS K 0102 46.3.1 Potassium peroxodisulfate decomposition method
Approx. 350 (W) × 500 (D) × 470 (H) mm AC100 V 50/60 Hz Reactor method Double plunger pump
Flow rate: 0.1~1.9 mL/min
Pressure resistance: 2.94 MPa (30 kgf/cm2)
Wetted parts: PTFE, ETFE, sapphire, ruby
Approximately 22 kg 880nm