Volatile organic substances (liquid samples) can be safely combusted in large quantities. The automatic sample combustion device AQF-2100V is ideal for pretreatment for measuring sulfur and halogens (fluorine, chlorine, bromine, and iodine) in liquid samples. Main business types: Fuel oil related, chemical quality control, environmental analysis Features

■Easy combustion

No combustion program required. By introducing the liquid sample into the device at low speed, anyone can easily and safely burn it.

■Can analyze samples with a concentration of 1ppm or less

Multiple combustion can be used to combust large amounts of liquid samples. ・Simultaneous measurement of halogen and sulfur is possible ・Achieve space saving

■Easy maintenance

The use of an open/close type electric furnace with a door placed in front of the combustion section makes it easy to install and remove the reaction tube for maintenance.

  • Product

    Sample combustion device (pretreatment) Automatic sample combustion device AQF-2100V

Share this product


70+ people viewing

Last viewed: 22 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Sample combustion device (pretreatment) Automatic sample combustion device AQF-2100V

Image Part Number Price (excluding tax) Power supply Plumbing Sample amount Weighing Combustion tube Combustion furnace Measurement target Dimensions/mass Absorption tube capacity Output Gas used
Sample combustion device (pretreatment) Automatic sample combustion device AQF-2100V-Part Number-AQF-2100V

AQF-2100V

Available upon quote

HF-210: AC 100/115/230/240V, 50/60Hz 1000VA
GA-210: AC 100/115/230/240V, 50/60Hz 50VA

Fluororesin tube, PEEK tube

Liquid sample: 5~100μl *Varies depending on the sample.

Loop weighing method

Made of high purity quartz

The electric furnace is divided into two parts and the temperature can be set individually. Maximum 1100℃

Liquid samples (gas injector is used for gas samples)

HF-210: Approx. 320 (W) × 430 (D) × 500 (H) mm, approximately 25 kg
GA-210: Approx. 250 (W) × 430 (D) × 500 ( (H) mm, approximately 22 kg

Made of Pyrex, content 10ml

Start signal for ion chromatography data processing (contact)

Argon: (99.98% purity or higher, 0.3±0.1MPa)
Oxygen: (purity 99.7% or more, 0.3±0.1MPa)

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Elemental Analyzers Products

Other products of Nittoseiko Analytech Co., Ltd

Reviews shown here are reviews of companies.


View more products of Nittoseiko Analytech Co., Ltd

About Company Handling This Product

Response Rate

100.0%


Response Time

64.4hrs

Company Overview

Nittoseiko Analytech Co., Ltd, founded in April 1979 and based in Yamato, Kanagawa, is a manufacturer and supp...

See More

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree