All Categories
History
Relatively Fast Response
Product
Hydrogen analyzer DH603Handling Company
Cosmo Trading Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | Power capacity | Sample size | Installation conditions | Furnace | Detection method | Minimum read | Size | Analytical accuracy (sample weight 1g) | Analysis range (sample weight 1g) | Reagents used | Gas used | Calibration | Gas flow rate | Option |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DH603 |
Available upon quote |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 35A, 23,600BTU/hr |
For non-diffusible hydrogen/maximum 23×102mm |
Temperature: 15℃~35℃ |
Resistance heating furnace (room temperature to 1,100℃) Temperature and heating rate can be programmed |
Thermal conductivity method |
Hydrogen: 0.001ppm |
Width: 57cm x Depth: 70cm x Height: 84cm |
Hydrogen: Standard deviation (s) 0.02ppm or RSD2% *Whichever is greater |
Hydrogen 0.1ppm~2,500ppm |
Magnesium perchlorate Mg (ClO4) 2 / Sodium hydroxide (NaOH) / Copper oxide (CuO) |
Carrier gas: Nitrogen purity 99.99%, 0.21MPa±10%, Nitrogen purity 99.99%, 0.41MPa±10% (when using diffusible hydrogen option), etc. |
Standard material/multipoint/gas dose |
3L/min (during operation) |
625-515 Upgrade Kit (DH603 → DH603DC) |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
24.3hrs
Company Review
4.0