Infrared microscope/imaging system Spotlight 400/400N-Spotlight 400 IR Imaging System
Infrared microscope/imaging system Spotlight 400/400N-Cosmo Trading Co., Ltd.

Infrared microscope/imaging system Spotlight 400/400N
Cosmo Trading Co., Ltd.


About This Product

The sample surface is scanned at high speed with an array detector, and information in the infrared region is expressed as a 2D sample or an image depending on the purpose. Imaging size can be freely set and data is acquired. Features ■Realizes ultra-high-speed IR imaging; imaging area can be freely changed We have adopted the patented SynchroScan interferometer synchronization system, which synchronizes the automatic stage of an infrared microscope and the linear MCT array detector for the first time in the world. This allows ultra-high-speed scanning. ■Compatible with imaging and micro IR measurement The patented Duet™ detector incorporates linear array elements and single elements coplanarly arranged into a dewar. In addition to imaging, transmission/reflection measurements, and microscopic ATR measurements are possible without switching detectors. Guarantees optimal performance for various applications. ■Detector pixel size can be selected from 1.56μm (ATR), 6.25μm, 25μm, 50μm Software control allows you to adjust the magnification to the detection element and specify the spatial resolution. There is no need for troublesome parts replacement. ■ATR imaging compatible Standard crystal supports wide area measurement up to 400µm x 400µm. The large type can measure a wide range of 750 µm × 750 µm to 400 µm × 1,000 µm. ■Imaging measurement possible down to low wave numbers Up to 650cm-1 as standard, with the industry's only broadband array (~580cm-1) and InGaAs detector available (optional). ■Spectrum search library (database) Library searching is one of the most important methods for analyzing samples measured by FTIR. Approximately 19,000 spectra of compounds, polymers, pharmaceuticals, etc. are included. ■NIR imaging measurement possible The Spotlight400N system is capable of imaging in the NIR region. - Ideal for obtaining NIR imaging in the pharmaceutical field, such as the distribution of ingredients in tablets and the distribution of active substances in bulk. - Can also be applied to non-destructive imaging of polymers and foods. ・No reflection peak is observed in solid sample reflection measurements in the near-infrared region. ・Advanced image processing is possible with HyperView software that supports principal component analysis (PCA), etc.

  • Product

    Infrared microscope/imaging system Spotlight 400/400N

Share this product


100+ people viewing

Last viewed: 4 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

2 Models of Infrared microscope/imaging system Spotlight 400/400N

Items marked with have different values depending on the model number.

Click on the part number for more information about each product

Product Image Part Number Price (excluding tax) ATR imaging area (large crystal) ATR imaging area (standard crystal) ATR imaging pixel size Auto (ATR) Auto control Imaging area Imaging measurement mode Imaging measurement wavenumber range (ATR) Imaging measurement wavenumber range (InGaAS) Imaging measurement wavenumber range (broadband MCT) Imaging measurement wavenumber range (transmission, reflection) Imaging pixel size Micro ATR (optional) Micro ATR (standard) Point measurement SN ratio Point measurement mode Point measurement wavenumber range (InGaAS) Point measurement wavenumber range (MCT) Size (W×H×D) Stage Stage size (standard stage) Stage size (surge stage, optional) Visible light source Visible observation Weight
Infrared microscope/imaging system Spotlight 400/400N-Part Number-Spotlight 400 IR Imaging System

Spotlight 400 IR Imaging System

Available upon quote Up to 750 µm × 750 µm, 400 µm × 1000 µm Up to 400 µm × 400 µm, 100 µm × 500 µm 6.25 µm × 6.25 µm, 1.56 µm × 1.56 µm Software controls crystal loading and unloading
Pressure monitor with software control
Line scan, mapping, XYZ stage, autofocus, autocorrection, aperture, transmission/reflection switching, infrared/visible switching, illumination brightness 100 µm × 100 µm ~ 50 mm × 50 mm arbitrarily variable Transmission/Reflection/ATR 4,500~680cm-1 7,800~4,000cm-1 (optional) 7,800~580cm-1 (optional) 7,800~650cm-1 50 µm × 50 µm, 25 µm × 25 µm, 6.25 µm × 6.25 µm 100µmφ conical Si crystal 100µmφ conical DLC-Ge crystal 100µmφ conical Ge crystal 12,000:1 (1 minute integration) Transmission/Reflection/ATR 10,000~4,000cm-1 7,800~600cm-1 Spotlight body: 340 x 470 x 570 mm
Spectrum 3 body: 520 × 300 × 600 mm
XYZ auto stage 75×50mm 160×60mm White LED CCD camera Spotlight: 32Kg
Spectrum 3: 34Kg
Infrared microscope/imaging system Spotlight 400/400N-Part Number-Spotlight 400N NIR Imaging System

Spotlight 400N NIR Imaging System

Available upon quote Up to 750 µm × 750 µm, 400 µm × 1000 µm Up to 400 µm × 400 µm, 100 µm × 500 µm 6.25 µm × 6.25 µm, 1.56 µm × 1.56 µm Software controls crystal loading and unloading
Pressure monitor with software control
Line scan, mapping, XYZ stage, autofocus, autocorrection, aperture, transmission/reflection switching, infrared/visible switching, illumination brightness 100 µm × 100 µm ~ 50 mm × 50 mm arbitrarily variable Transmission/Reflection/ATR 4,500~680cm-1 7,800~4,000cm-1 (optional) 7,800~580cm-1 (optional) 7,800~650cm-1 50 µm × 50 µm, 25 µm × 25 µm, 6.25 µm × 6.25 µm 100µmφ conical Si crystal 100µmφ conical DLC-Ge crystal 100µmφ conical Ge crystal 12,000:1 (1 minute integration) Transmission/Reflection/ATR 10,000~4,000cm-1 7,800~600cm-1 Spotlight body: 340 x 470 x 570 mm
Spectrum 3 body: 520 × 300 × 600 mm
XYZ auto stage 75×50mm 160×60mm White LED CCD camera Spotlight: 32Kg
Spectrum 3: 34Kg

Click on the part number for more information about each product