Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50-Model-4001B-50
Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50-Showa Sokki Co., Ltd.

Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50
Showa Sokki Co., Ltd.


About This Product

■Features

・Charge amplifier for piezo force sensors and pressure sensors. ・Unlike general charge amplifiers, this charge amplifier can measure from almost DC. ・The discharge time constant can be switched between LONG/SHORT, and the low-frequency response characteristic can be selected.

■Charge amplifier Model-4001B-50 is suitable for the following applications - “Force” measurement with a wide dynamic range

・Measurement of impact “force” ・Measurement of engine combustion pressure ・Measurement of compressor pressure, etc.

  • Product

    Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50

Share this product


20+ people viewing

Last viewed: 55 minutes ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50

Image Part Number Price (excluding tax) Power supply Mass Linearity Measuring range Temperature, humidity range Maximum input charge Maximum drift Discharge time constant (TC) Frequency characteristics Output current Output voltage Output calibration signal Output resistance Input equivalent noise Low frequency cutoff Range accuracy Reset error Reset switching time Over load
Charge amplifier and power supply for sensors with built-in amplifier Charge amplifier for force and pressure sensors Model-4001B-50-Part Number-Model-4001B-50

Model-4001B-50

Available upon quote

Charge amplifier ±15VDC, 20mA or less
AC100V±10V when using AC adapter
When using power supply unit AC100V±10V

Charge amplifier MODEL-4035-50310g/CH
Power supply MODEL-4001-10A 800g
Case for 6CH MODEL-4001-26: 1,300g

Within ±0.1%

Detector sensitivity 1.00~9.99pC/UNIT: 10, 20, 50, 100, 200, 500, 1,000UNIT/VOLT

0~50℃, 80%RH or less

±100,000pC (slew rate 5,000pC/μs)

Can be adjusted to below ±0.03pC/s with DC BAL (after 20 minutes of warm-up)

SHORT range x 0.1sec (feedback resistance 109Ω)
LONG range x 1,000sec (feedback resistance 1,013Ω or more)

Range 100~1,000DC~50kHz within 5%
Range 50 DC~40kHz within 5%
Range 20 DC~20kHz within 5%
Range 10 DC~10kHz within 5%

Maximum ±5mA

Max ±10V

Supplied from power supply MODEL-4001-10A 0~10VDC variable
With CAL/OP changeover switch (back)

100Ω

0.1pCrms or less

0.16/ Discharge time constant (TC) Hz (-3dB)

Within 1.5% (at 80Hz)

±0.5pC or less

500msec or more

0~50℃, 80%RH or less

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Pressure Sensors Products

Other products of Showa Sokki Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Showa Sokki Co., Ltd.

About Company Handling This Product

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree