Organic analysis Fully automatic moisture analyzer TGM800 typeHandling Company
LECO Japan LLCCategories
Product Image | Part Number | Price (excluding tax) | Analysis accuracy | Balance reading limit | Display | Gas flow | Gas used | Heating rate | Installation conditions | Number of samples | Oven temperature | Power capacity | Sample amount | Size | Temperature control | Ventilation | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
TGM800 |
Available upon quote | 0.02% RSD (1s / inert sample) | 0.0001g | 10.1 inch touch screen type color LCD | 0~5 L/min | Air (no moisture or oil) or nitrogen 99.5%, 0.24 MPa ±10% | 25℃/min (room temperature to 104℃) |
Room temperature: 15℃~35℃, relative humidity: 20%~80% (non-condensing) Avoid wind, vibration and direct sunlight |
16 (+1 reference) 10 (+1 reference) |
Room temperature (minimum) | 115 V (+10 %), 50/60 Hz, single phase 10 A (max), 1,500 BTU/hr* |
1.5 inch crucible: 1 g (typical) 2.4 inch crucible: 3 g (typical) |
Width: 39 cm x Depth: 74 cm x Height: 41 cm |
50℃~175℃ Accuracy: 2% of set temperature or ±2℃ *Whichever is greater Stability: 2% of set temperature or ±2℃ *Whichever is greater |
Optionally, 1 inch exhaust duct can be installed within 3 m. | Approximately 34 kg |