FLC-103: Oxygen flow meter flow measurement/diagnosis-FLC-103
FLC-103: Oxygen flow meter flow measurement/diagnosis-Kowa Keiso Co., Ltd.

FLC-103: Oxygen flow meter flow measurement/diagnosis
Kowa Keiso Co., Ltd.

🏥 Medical Industry Use

About This Product

■Is the oxygen flow rate really correct? Oxygen flowmeters are used almost every day in medical institutions. In the case of a flow type, due to its structure, wear of the float and deterioration of the packing may cause If used for a long period of time, the accuracy will decrease. Also, if the dial type is used, check whether the set number of oxygen is being administered. Since I don't actually know, If the product has been damaged due to bumping, etc., or due to deterioration over time, etc. The dial value may not match the actual oxygen flow rate. For the reasons mentioned above, oxygen flowmeters are also devices that require periodic inspection. If a medical institution attempts to check the accuracy of an oxygen flow meter by itself, Various equipment (pressure reducing valves, tubes, filters, flow meters, etc.) This may result in poor testing efficiency, and depending on the scale of the medical institution. We may have hundreds of devices or more, Therefore, there is a problem that self-inspection is difficult. The oxygen flow meter checker includes the equipment necessary to inspect oxygen flow meters. This product is compactly packaged. In addition, it takes less time to measure the oxygen flow rate and has a judgment function, so It has the feature that anyone can easily operate it. This is the perfect product for those who are considering inspecting oxygen flow meters. ■Features ・You can easily measure the flow rate of an oxygen flow meter. Comes with automatic judgment function You can see OK/NG at a glance. *Standard values ​​vary depending on the manufacturer, so please check the attached document. -The display is digital, so anyone can easily read the flow rate. -Small and lightweight. In addition, it has a hook and can be attached to the hospital outlet section. Can be used by hanging. Anyone can test efficiently and easily. ・Can inspect from minute flow rate (0.5L/min) to medium flow rate (15L/min). ・Since it is battery operated, it can be used in places where there is no outlet. You can take it anywhere for inspection. ■Outlet attachment (optional) Can be attached to the outlet. Used when checking outlet pressure. Comes with digital pressure gauge. *Since this is a made-to-order item, the delivery time will be long. ■Gas cylinder attachment (optional) Can be attached to screw-in type cylinders. Flow meter with pressure reducing valve Used when checking the pressure gauge. Comes with digital pressure gauge. *Since this is a made-to-order item, the delivery time will be long.

  • Product

    FLC-103: Oxygen flow meter flow measurement/diagnosis
  • Usage Scenarios

    Medical Industry Use

Share this product


240+ people viewing

Last viewed: 15 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of FLC-103: Oxygen flow meter flow measurement/diagnosis

Product Image Part Number Price (excluding tax) Accessories Dimensions/mass Judgment criteria Judgment result display method Measurement channel Measurement method Measurement target Measuring range Option Power supply Purpose Resolution Usage environment/storage environment Warranty period
FLC-103: Oxygen flow meter flow measurement/diagnosis-Part Number-FLC-103

FLC-103

Available upon quote AA batteries x 4 for operation check
1 x tube with flare adapter
Tube x 1
Protective cap x 2
W177.8 x H69 x D93.2 mm / approx. 500g 0.5L min, 1L/min: Passed within ±30% error
2L/min: Pass within ±20% error
5L/min, 10L/min, 15L/min: Pass with error within ±10%
LED display (pass: lit green, fail: lit red) 1 channel By semiconductor mass flow sensor ①Pin type or Schrader type float type oxygen flow meter
②Pin type or Schrader type dial type oxygen flow meter
0.0L/min~20L/min FLC-OA1: Outlet attachment
FLC-BA1: Gas cylinder attachment
AA alkaline batteries x 4 (DC6V) Flow measurement using a medical oxygen flow meter (fluid is oxygen) ±0.15%F.S. or ±3%R.O., whichever is greater No condensation, oil smoke, corrosive gas, conductive dust, etc.
No high voltage equipment nearby
Room temperature 15~35℃
1 year from purchase