All Categories

History

Research and development using diffraction gratings - Optical spectrum analyzer "OSA20-OSA20
Research and development using diffraction gratings - Optical spectrum analyzer "OSA20-Seven Six Co., Ltd.

Research and development using diffraction gratings - Optical spectrum analyzer "OSA20
Seven Six Co., Ltd.

Seven Six Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

52.0hours


About This Product

EXFO's OSA20 is an optical spectrum analyzer for research and development sites that uses a diffraction grating, and features high-speed sweep, high resolution, and high wavelength accuracy.

■Features

・High-speed sweep up to 2,000nm/s ・High resolution bandwidth up to 20pm (adjustable from 50 to 2,000pm) ・High wavelength accuracy ±10 pm@1,500 to 1,640nm ±25 pm@1,250 to 1,700nm ・Wide wavelength range 1,250 to 1,700nm ・Equipped with 8 measurement modes OSA/WDM/MML/RLT/OFA/SML/BBS/PCT ・Equipped with 12inch touch panel display ・Application: Research and development use

■Notes

*a. User calibration will be performed after 1 hour warm-up time. *b. Native 17-24pm (1,500-1,620nm) (excluding sensitivity -55dBm), 17-26pm (1,250-1,700nm) *c. Adjustable resolution bandwidth is calculated from native bandwidth *d. Adjustable in 0.1GHz steps in the range of 6 to 400GHz (Unit: THz) *e. Uses a 1,523nm HeNe laser with a span of ±2nm. *f. 1,523nm laser (±50nm span excluding ±2nm near the peak) *g. Noise level for 99% of all data points at 1,520-1,620 nm *h. Typical value *i. At 18-28℃, all sensitivity settings except -55 dBm and burst sensitivity of ±0.6 dB. *j. Measured values ​​at 1,310nm and 1,500nm. except for -55 dBm and ±0.3 dB at burst sensitivity *k. Excluding water absorption line, ±0.35 dB at -55 dBm and burst sensitivity, over all sensitivity settings from 18°C ​​to 28°C. *l. Sweep period / 100 nm, sensitivity -60 dBm, center wavelength 1,475 nm

  • Product

    Research and development using diffraction gratings - Optical spectrum analyzer "OSA20

Share this product


190+ people viewing

Last viewed: 8 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Research and development using diffraction gratings - Optical spectrum analyzer "OSA20

Image Part Number Price (excluding tax) Wavelength Accuracy 2σ *a Wavelength Linearity 2σ *a Wavelength Reference source Wavelength Reproducibility 2σ Wavelength Spectral range Wavelength Span range Wavelength Sampling resolution Wavelength Sampling point Optical interface return loss Optical interface input/output connector Optical interface optical input Optical interface calibration output Monochromator Stray light suppression rate *f Monochromator resolution bandwidth *b, c, d Monochromator Dynamic Range (ORR) *e Sweep speed sensitivity *g Sweep speed Sweep cycle/100 nm *l Sweep speed Sampling rate OpticalPower refracting power Level accuracy *a, i OpticalPower refracting power Level wavelength flatness *k OpticalPower refractive power Level linearity OpticalPower power level sampling OpticalPower Refractive Power Total Safe Power Optical powerRefractive powerInput power for each channel Optical Power Refractive power Level sensitivity
Research and development using diffraction gratings - Optical spectrum analyzer "OSA20-Part Number-OSA20

OSA20

Available upon quote

±10 pm over 1,500-1,640 nm, ±25 pm over full range

±6 pm over 1,500-1,640 nm, ±20 pm over full range

Built-in ELED (safety class 1) + Acetylene cell (user calibration by patchcord)

±2 pm / 0.25 GHz at 1,550 nm

1,250-1,700 nm / 239.834-176.349 THz

0.5 nm to full range (450 nm)

2pm / 0.25GHz at 1,550nm

251 (span of 0.5 nm) to 225,001 (span of 450 nm)

38 dB (> 42 dB typ.) at 1,310 nm and at 1,550 nm (APC connector

FC/APC or FC/PC or SC/APC or SC/PC

SMF-28 type fiber

Wavelength reference (see above)

≥73dB

20 pm native (20 pm = 2.5 GHz), adjustable over 50–2,000 pm with 1 pm step

≥ 30 dB (> 35 dB typ.) beyond ±50 pm from peak,≥ 50 dB (> 55 dB typ.) beyond ±100 pm from peak, etc.

-55 dBm at 2,000 nm/s to -75 dBm at 2 nm/s

300ms typ.

1MHz typical

±0.4 dB at 1,310 nm and 1,550 nm

±0.15 dB over 1,500-1,640 nm, ±0.25 dB over 1,260-1,680 nm

±0.07 dB over the full range (input level —50 to +3 dBm)

±0.01 dB over —60 to +20 dBm

≤25dBm

≤20dBm

Single scan: High (0.5 nm/s): < —76 dBm (—78 dBm typ.)

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Spectrum Analyzers Products

Other products of Seven Six Co., Ltd.