All Categories

History

Multi -use data logger MIDI LOGGER GLT400-GLT400
Multi -use data logger MIDI LOGGER GLT400-Graphtech Co., Ltd.

Multi -use data logger MIDI LOGGER GLT400
Graphtech Co., Ltd.

Graphtech Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

19.7hours

Relatively Fast Response


About This Product

■ Multi -use multi -function input compatible GLT400

In addition to the use of research and development sites, the GLT400 has expanded the department to the production technology and production site, incorporated it into the control panel and equipment, acquired data from the PLC with MODBUS/TCP, and as a PC logger. It is a multi -use data logger that can be used as a standalone measurement and a GL840 sliding unit such as use. Three types of analog input terminals can be selected according to the lineup and input form, and can be expanded from standard 20ch to up to 200ch. We offer connection procedures with PLCs using modbus/TCP. <How to use GLT400>

■ Used as a standalone

・ Incorporated into the control panel or device ・ Modbus linkage as PLC I/O ・ Can be used as a PC logger

■ Used as a sliding unit of GL840

・ Wireless LAN connection ・ Ethernet connection ・ Wireless LAN & Ethernet connection <Multi -function input support> Voltage, temperature, and humidity can be selected for each input channel

■ I want to measure the voltage

Equipped with abundant range from 20MV to 100V

■ I want to measure the temperature

・ Thermal electricity vs.: K, J, T, R, S, B, N, C (former W: WRE5-26) ・ Temperature resistance body: PT100, JPT100, PT1,000 (IEC751)

■ I want to measure the current

・ Prepare a shunt resistance of 250Ω as an option ・ Equipped with a dedicated 1-5V range in the measurement range

■ I want to measure humidity

Exclusive humidity sensor is prepared as an option, and sensor power supply is supplied from the main unit, and a dedicated humidity range is installed.

■ Equipped with digital I/O

Digital I/O can be used by connecting the optional input cable.

■ Logic/pulse input (either 4 points)

・ Logic: H.L judgment by voltage contact point (A contact, B contact, NO, NC), open collector, voltage input ・ Pulse: You can select for each channel from 3 modes (rotation speed / accumulation / instant)

■ Alarm output (4 points) & external trigger signal input (1 point)

・ Measurement can be started and stopped at an external electrical signal. ・ Set an abnormal level and output port (1 to 4) for each analog input channel, and the signal output is performed by abnormal detection.

  • Product

    Multi -use data logger MIDI LOGGER GLT400

Share this product


320+ people viewing


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Multi -use data logger MIDI LOGGER GLT400

Image Part Number Price (excluding tax) Number of analog terminal units Analog terminal unit type Sampling cycle Repeat trigger Trigger type Trigger setting Alarm judgment type Input type Input specifications Output type Output specifications Measuring mode Maximum number of pulse input PC I/F function Ethernet function USB function Real -time transfer speed Body memory External memory slot 1 file maximum capacity Memory content usage environment AC adapter DC input USB power supply power consumption External dimensions (W × D × H) (approx.) Mass (weight) (approx.) Earthquake resistance
Multi -use data logger MIDI LOGGER GLT400-Part Number-GLT400

GLT400

$ 650.00〜

Up to 10 units (200ch)

Standard terminals, high -pressure high -precision terminals, screwless terminals

10ms/ch ~ 1H (10ms to 50ms only voltage, CH mathematical limit), external (only at Stand Alone)

OFF / ON

Start: Start data recording due to trigger, stop: Stop data due to trigger.

Start: OFF, Level, Alarm, External, Time, Day, Regular time, Stop: OFF / Level / Alarm / External / Time / Day

Analog judgment: Standing up (↑), falling down (↓), within the range, outside the range, logic judgment: Pattern, pulse judgment: Standing up (↑), falling (↓), within the range

Trigger input (1ch) or external sampling input (1ch), logic input (4ch) or pulse input (4ch) (only at Stand Alone)

Input voltage range: 0 ~ + 30V (single -line contact input), input signal: non -voltage contact point (contact point A, B contact, NC), open collector, voltage input

Alarm output: 4ch (only 1ch at Remote) * Switch for logic, pulse, trigger and external sampling

Output format: Open collector output (5V pull -up resistance 10kΩ)

Rotation mode/accumulation mode/instant mode

Maximum input frequency: 50kHz, maximum count: 50kc/sampling (16bit counter)

Ethernet (10BASE-T / 100BASE-TX), USB2.0 (high speed), wireless LAN (when optional)

Data transfer to PCs, body control from PC, network connection as a sliding unit of GL840

Web server function, FTP server function, FTP client function, NTP client function, DHCP client function, DHC server function, MODBUS/TCP communication function

USB drive mode (body memory or recorded data file in SD memory card PC transfer / deleted)

The fastest 10ms/1ch (the transfer speed varies depending on the number of channels)

Approximately 8GB

SD memory card slot: 1 (SDHC compatible, up to about 32GB memory can be used)

1 file 2GB

Setting conditions / measurement data of the main unit

-20-60 ° C (attached AC adapter and B-565 for 0 to 45 ° C), 5-85 % RH (without condensation) (if USB PD is powered, it depends on the specification of the power supply).

AC 100-240V / 50-60Hz

DC 8.5-24V (up to 26.4V) (Optional B-514 DC drive cable is required)

External USB PD compatible battery (standard: USB Power Delivery 2.0 or later, output: 5V 2A (10W))

24VA or less (using AC adapter, AC100V)

When the standard terminal is installed/when the screwless terminal is installed: 187.5 × 183 × 65.5 mm (including protrusions), high -pressure and high -precision terminal installation: 187.5 × 183 × 73.4 mm (not included)

When the standard terminal is installed: 1,090g (not included in AC Adabuta), when the screwless terminal is installed: 1,020g (not included in AC advertising), when high -pressure and high accuracy terminals are installed: 1,120g (not included in AC Adabuta)

Automotive parts 1 type 1 type A equivalent

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Data Loggers Products

Other products of Graphtech Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Graphtech Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

19.7hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree