About This Product

The lightweight and compact aluminum body is protected by a removable rubber protective cover, making it perfect for use in harsh environmental conditions. Suitable for bright outdoors. Low reflection transmissive type, with backlight. Adopts an easy-to-use operation/display menu screen based on Windows specifications. Displays parameters in a list format that is easy to see and search. You can freely select any 5 items out of 64 items and set them to F1-F5. Easy to replace Most important items have dedicated keys.

■Features

・Uses lithium battery, operates for 13 hours, energy saving mode, and safely charges automatically with internal charging circuit processor ・Color setting of measurement screen and menu screen possible ・Parameters, Help text, etc. can be edited on a computer, transferred to the main unit, and saved. - Abundant setting assist functions: For example, you can easily set the delay material length correction of the probe. ・USB printer interface - Upgradable from external PC via USB ・Pulse repetition frequency (PRF): 1500 Hz ・Real-time hardware gate output (2 gates) ・Synchronization mode: Internal/External (in/out) /S echo synchronization monitor gate

■Number of monitor gates

2

■Response speed

Both gates 1 and 2 respond to pulse repetition frequency (Max.1,500Hz)

■Operating mode

Normal, Invert, OFF

■Gate setting range

・Gate delay: 0–4,000mm ・Gate width: 0.1–3,000mm Both in 0.1mm steps

■Continuous defect number setting range

0–250 continuous pulses (common to both gates)

■Relay output Go/NoGo (common to both gates)

・TTL level (5V), low:active, ZA=100Ω Response accuracy: +/-0.5% echo height ・Switching hysteresis:<0.5% echo height ・Output hold time: 12ms

■Analog output

0-2.5V (0-100% proportional to height) Gate 1/Gate 2

■Distance and thickness measurement

Transmission - Gate 1/Gate 1 - Gate 2

■Audible alarm buzzer

KHzD 1 each, total 2 (front panel)

■Visible warning lamp

1 LED each, 2 total (front panel) Input/Output*Options

■USB interface*

・For PC connection, separately sold software ECOM90 is required. ・Data can be output and saved to a PC

■Synchronous input/output

TTL level (5V), falling sync, sync threshold: approx. 2V

■Monitor output

VGA

  • Product

    Ultrasonic flaw detector echograph

Share this product


70+ people viewing

Last viewed: 22 minutes ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Ultrasonic flaw detector echograph

Image Part Number Price (excluding tax)
Ultrasonic flaw detector echograph-Part Number-1090 BASIC

1090 BASIC

Available upon quote

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Ultrasonic Flaw Detectors Products

Other products of Nihon Matec Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Nihon Matec Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

1.5hrs

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2025 Metoree