Product
AIR Atmospheric/Gas Measuring Instrument Oxygen Automatic Measuring Instrument EMG-88 LiteHandling Company
Anatec Yanaco Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Automatic calibration | External dimensions | Indication error | Mass | Measurement method | Measuring range | Output signal | Paint color | Power consumption | Power supply | Proofreading | Repeatability | Response time | Sample amount | Span drift | Zero drift |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
EMG-88 Lite |
Available upon quote |
Calibration cycle: 1 to 999 hours (can be set arbitrarily in 1 hour steps) Number of calibration components: 1 component (O₂ zero, span calibration) Calibration accuracy: Within ±0.5% of full scale (both zero and span) |
(W) 400 × (D) 300 × (H) 450mm (excluding protrusions such as handles) | ±2% of full scale | Approximately 17kg | Magnetic force type | 0~10,25% (tested) | DC0~1V (concentration signal/non-insulator output) | Munsell 5Y7.5/1 semi-gloss | Approximately 200VA | AC100V±10V 50/60Hz shared |
Zero: N₂ Span: O₂/N₂ (80~100% of full scale) |
±1% of full scale | 90% response time from instrument entrance: within 60 seconds (standard) | Approximately 1L/min | ±2%/week of full scale | ±2%/week of full scale |