All Categories
History
Product
Dual output digital pressure controller 7252, 7252Handling Company
Fluke CalibrationCategories
Image | Part Number | Price (excluding tax) | Power supply | Weight | Communication | Negative gauge accuracy (optional) | Display resolution | Accuracy Model 7252 | Vacuum pressure reference (optional) | Humidity environment | Temperature environment | Proofreading | Dimensions (H × W × D) | Stability | Atmospheric pressure reference (optional) | Pressure medium | Control response | Control stability | Control parameters | Menu and functions | Driver | Display | Test port and supply pressure port | Warm up time |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
7252 |
Available upon quote |
90~260VAC, 50/60Hz, 150W |
Series 7252/7252i: 9 kg |
RS232 and IEEE-488, SCPI syntax |
0.003% of positive full scale |
User selectable, up to 1:1,000,000 |
Range 5 to 2500 psi (400 mbar to 172 bar): 0.003% of full scale |
0.0002 psi (0.014 mbar) maximum error/year |
Relative humidity 5 to 95%, non-condensing |
Usage: 18°C ~ 36°C (64°F ~ 97°F) |
A National Institute of Standards and Technology (NIST) traceable calibration report is provided. NVLAP certified calibration available. |
All versions: 178 mm x 419 mm x 483 mm |
0.0075% of reading/year |
Up to 0.014 kPa in 1 year |
High purity nitrogen or dust free dry air |
Less than 15 seconds with zero overshoot to 15 in3 (245 cc) volume in 10% increments |
Active mode: 0.001% of each range |
Volume: |
The Series 7252 can display menus and functions in English, Japanese, French, Spanish, Chinese, Italian, and German. |
LabVIEW driver |
TFT, VGA, active matrix, 6.4 in (163 mm), resolution 640 × 480, 65,000 colors |
1/4 in NPT female |
2 ~ 3 hours, (leave with electricity on) |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.