The LB6701-H10 is an ionization chamber probe used for dose rate measurement in various radiation protection fields and as a dose rate probe for photon radiation in the health physics field. The measured radiation dose is the ambient dose equivalent H* (10) and the ambient dose equivalent rate Ḣ* (10). The ionization chamber is made of aluminum and filled with 10bar nitrogen gas, making it radiation resistant up to 106 Gy.

■There are three versions of the ionization chamber probe:

・LB6701L-H10 (Low dose rate:10 µSv/h ~ 10 Sv/h) ・LB6701M-H10 (Middle dose rate:0.1 mSv/h ~ 100 Sv/h) ・LB6701H-H10 (High dose rate: 1 mSv/h ~ 1000 Sv/h)

  • Product

    Ionizing radiation dose rate monitor LB6701

Share this product


50+ people viewing


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

3 Models of Ionizing radiation dose rate monitor LB6701

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Measuring range Operating environment Relative humidity Operating environment Temperature Operating environment Storage temperature Output pulse intermediate frequency converter polarity Output pulse intermediate frequency converter amplitude Output pulse intermediate frequency converter frequency range Output pulse Intermediate frequency converter Pulse width Protection class High voltage Size Ionization chamber Size I-F converter connection box Calibration factor Energy range
Ionizing radiation dose rate monitor LB6701-Part Number-LB6701L-H10

LB6701L-H10

Available upon quote

10 μSv/h – 10 Sv/h

20~80%

0~+50°C

Up to 60°C

Positive

0,1Hz – 100kHz

2 – 5 µs

IP65

1000 Volts

80mm Ø x 268mm

240x160x90mm

0,1 mSv/h per cps

45 keV – 1,3 MeV with regard to Cs-137 and 0°

Ionizing radiation dose rate monitor LB6701-Part Number-LB6701M-H10

LB6701M-H10

Available upon quote

0,1 mSv/h – 100 Sv/h

20~80%

0~+50°C

Up to 60°C

Positive

0,1Hz – 100kHz

2 – 5 µs

IP65

1000 Volts

80mm Ø x 268mm

240x160x90mm

1 mSv/h per cps

45 keV – 1,3 MeV with regard to Cs-137 and 0°

Ionizing radiation dose rate monitor LB6701-Part Number-LB6701H-H10

LB6701H-H10

Available upon quote

1mSv/h – 1000Sv/h

20~80%

0~+50°C

Up to 60°C

Positive

0,1Hz – 100kHz

2 – 5 µs

IP65

1000 Volts

80mm Ø x 268mm

240x160x90mm

10 mSv/h per cps

45 keV – 1,3 MeV with regard to Cs-137 and 0°

Click on the part number for more information about each product

Other products of Berthold Japan Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Berthold Japan Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

36.4hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2025 Metoree