Mitochondrial metabolism analysis system/cell flux analyzer OROBOROS O2k-FluoRespirometerHandling Company
Berthold Japan Co., Ltd.Image | Part Number | Price (excluding tax) | Oxygen signal Noise during anoxia | Oxygen signal time constant | Oxygen signal Noise due to atmospheric saturation | Oxygen Signal O2 Linearity Range | System operating environment Temperature stability | System operating environment Recommended PC | System operating environment Operating temperature | System operating environment Size/Weight |
---|---|---|---|---|---|---|---|---|---|---|
![]() |
OROBOROS O2k-FluoRespirometer |
Available upon quote |
< ± 0.02 µM (SD, temperature 37 °C, 100 data points recorded at 0.2 second intervals) / |
<4 seconds at 37℃ (<3 seconds typical) |
< ± 0.1 µM O2 (SD, temperature 37 °C, 100 data points recorded at 0.2 second intervals) / |
0 – 100 µM |
±0.002℃ |
OS Windows XP, CPU Intel-Core-2 or higher, memory 2 GB RAM or |
4 – 47 °C (continuous monitoring) |
310 mm (W) × 450 mm (D) × 250 mm (H) / 13.5 kg (main unit) |