DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment-DSCN-101A
DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment-Kowa Keiso Co., Ltd.

DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment
Kowa Keiso Co., Ltd.

🏥 Medical Industry Use 💻 Electronics & Electrical Use

Kowa Keiso Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

62.0hours


About This Product

■Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment

・Power cord resistance diagnosis ・Grounding resistance diagnosis for equipment whose power cord cannot be removed ・3P table tap resistance diagnosis

■Why is it necessary to measure the resistance of the power cord?

・Poor contact or disconnection may occur due to insertion/removal, twisting, or deterioration over time. ・If the resistance value of the ground line is large, there is a risk of electric shock. JST T0601 (corresponding international standard IEC 6060-1) <Medical electrical equipment - Part 1: Excerpt from general safety requirements> Excerpt <Impedance and current carrying capacity> For ME equipment equipped with an equipment power socket, the protective grounding blade of the equipment power socket The impedance to all protective earthed parts shall not exceed 0.1Ω. ↓ In summary, the resistance of the ground line should not exceed 0.1Ω. The purpose of this is to prevent microshock (tolerance: 0.01mA*) *1/10 of 0.1mA that causes ventricular fibrillation when in direct contact with the heart 100 power cables actually used at a medical institution As a result of testing the grounding wire resistance value, 60% of the cables had a resistance of 0.11Ω or more. There is also data.

■Advantages of using a power cable checker

At some medical institutions, the number of medical devices returned each day ranges from 30 to 50. Abnormalities in appearance are easy to notice, but the resistance value of the power cord needs to be measured. Attempting to measure this using a commercially available tester would require a great deal of effort. Even if you try to measure the resistance value with a tester, the following problems will occur. - Unstable contact between probe and terminal ・It is necessary to measure the ground wire E, power wire L, and power wire N three times in total. - Inaccurate ground wire resistance measurement ・It takes time from preparation to completion of measurement ・Can only be measured by someone with specialized knowledge By using a power cable checker Problems with tester measurement can be solved at once. ・Measurements can be made in a stable position due to the cable plug-in type product design. ・With the cable plug-in type, two power wires and one ground wire can be measured at one time. ・Accurately measure ground wire resistance by adopting a high-precision 4-terminal measurement method ・Resistance measurement completed in about 10 seconds - Anyone can use it, just plug in the cable and press the measurement switch. - Battery operated so you can use it anywhere

  • Product

    DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment
  • Usage Scenarios

    Medical Industry Use / Electronics & Electrical Use

Share this product


260+ people viewing


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment

Image Part Number Price (excluding tax) Power supply Accuracy Purpose Measurement voltage Measurement method Measurement target Measuring range Number of measurement channels Dimensions Mass Frequency Judgment result display method Judgment criteria Power cord Judgment criteria External terminal Resolution Warranty period Usage environment Storage environment Accessories
DSCN-101A: Measurement and diagnosis of ground wire resistance and power cord resistance of ME equipment-Part Number-DSCN-101A

DSCN-101A

Available upon quote

9V 006P type dry battery x1 (DC9V)

±5% of reading

Measuring the resistance of equipment power cords

5V

Resistance measurement using AC 4-terminal method

・3P type plug power cord with grounding terminal
・3P type plug power cord that cannot be removed from the device

Up to 10Ω

1 channel

W177.8xH69xD93.2mm/approx. 490g (including battery)

AC 1kHz

LED display (normal: lit green, abnormal: lit red, unconnected: lit red)

Grounding electrode resistance value is 0.1Ω or less: OK, over 0.1Ω: NG
Ground side pole and voltage side pole resistance values ​​are 5Ω: OK, exceeding 5Ω: NG

Grounding wire pole resistance value is 0.2Ω or less: OK, exceeding 0.2Ω: NG
*The ground electrode side and voltage side electrode are not measured.

0.001Ω

1 year from purchase

No condensation, oil smoke, corrosive gas, conductive dust, etc.
No high voltage equipment nearby
Room temperature 5℃~40℃

9V 006P battery x1, test cord x1, external cord x1, cotton swab x1

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Continuity Checkers Products

Other products of Kowa Keiso Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Kowa Keiso Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

62.0hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2025 Metoree