Product
Oxygen Nitrogen Hydrogen Analyzer ONH836 SeriesHandling Company
Cosmo Trading Co., Ltd.Categories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Analysis range (sample weight 1g) | Analysis time | Analytical accuracy (sample weight 1g) | Calibration | Detection method | Gas flow rate | Gas options | Gas used | Internal cooling water | Operating environment | Part number | Power capacity | Reagent | Sample weight | Size | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ONH836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
ON836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
OH836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
NH836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
O836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
N836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) | |
H836 |
Available upon quote |
Oxygen: 0.05ppm~5.0% Nitrogen (carrier gas helium): 0.05ppm~3.0% Nitrogen (carrier gas argon): 0.2ppm~3.0% Hydrogen: 0.1ppm~2,500ppm |
Oxygen [carrier gas helium]: 85 seconds, [carrier gas argon]: 95 seconds, etc. (Including outgassing, purge, and analysis delay) |
Oxygen: Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is greater Nitrogen (carrier gas helium): Standard deviation (s) 0.025ppm or RSD0.3% *Whichever is larger, etc. |
Standard sample (1 point or multiple points), manual: gas dose |
Oxygen, hydrogen: non-dispersive infrared absorption method Nitrogen: thermal conductivity method |
Carrier: 490cc/min Drive: 280cc/min |
Gas dose: carbon dioxide (99.99%) 0.14MPa, nitrogen (99.99%) 0.14MPa |
Carrier: Helium (99.99%), 0.15MPa or Argon (99.999%), 0.15MPa Drive: Dry nitrogen, 0.28MPa (contains no moisture or oil) |
Distilled water 3.2 L |
Temperature: 15℃~35℃ Relative humidity: 20%~80% (non-condensing) |
[ONH836-MC] Oxygen/nitrogen/hydrogen analyzer, mount type monitor [ONH836-HMC] Oxygen/nitrogen/hydrogen analyzers, auto cleaners, mount type monitors, etc. |
Single phase 200V~ (+10/-15%: maximum output, boosted to 230V), 50/60Hz, 30A, 12,500BTU/hr *Average when operating with normal settings |
・Magnesium perchlorate ・Copper oxide, reduced copper ・Sodium hydroxide (coated on an inert substrate) ・Oxygen water removal agent (OMI tube) |
Usually 1g |
Width: 71cm Depth: with touch screen monitor (85cm) No touch screen monitor (76cm) Height: When using road head cover lift (100cm), Normally (91.5cm) |
Body weight 186kg (without touch screen monitor) |