Product
Surface temperature sensor calibrator ST-CAL1Handling Company
RKC INSTRUMENTCategories
Product Image | Part Number | Price (excluding tax) | Absolute humidity | Accessories | Accuracy | Communication method | Communication protocol | Communication speed | Control method | Cooling time | Heating time | Insulation resistance calibration furnace unit | Insulation resistance temperature control unit | Mass Calibration furnace unit | Mass Heater cable | Mass sensor cable | Mass temperature control unit | Operating environment conditions Ambient humidity | Operating environment conditions Ambient temperature | Overheating prevention function | Power consumption/current (at maximum load) | Power supply voltage | Sampling period | Set temperature range | Stability time | Surface temperature distribution | Temperature sensor | Temperature stability | Withstand voltage calibration furnace unit | Withstand voltage temperature control unit |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ST-CAL1 |
Available upon quote | MAX 29g/m3 Dry air at 101.3kPa | Heater cable, sensor cable, test report, instruction manual |
-0.6~+0.2℃ (≦100℃) -0.9~+0.2℃ (≦200℃) -1.2~+0.2℃≦300℃) |
RS-485 | RKC standard communication | 2,400/4,800/9,600/19,200/38,400bps | Brilliant PID control (phase control) |
Approx. 10 minutes (300→250℃), Approx. 15 minutes (300→200℃), Approx. 25 minutes (300→150℃), Approx. 45 minutes (300→100℃), Approx. 85 minutes (300→50℃) *Temperature drop time in natural cooling (ambient temperature 23℃) |
Approx. 2 minutes (23→50℃), Approx. 3 minutes (23→100℃), Approx. 4 minutes (23→150℃), Approx. 5 minutes (23→200℃), Approx. 7 minutes (23→250℃), Approximately 8 minutes (23→300℃) |
Sensor connector: 100MΩ or more between ground terminals (DC 100V) Heater connector: 20MΩ or more between ground terminals (DC 500V) |
Sensor connector: 20MΩ or more between ground terminals (DC 500V) Heater connector: 20MΩ or more between ground terminals (DC 500V) 20MΩ or more between power terminal and ground terminal (DC 500V) |
Approximately 4.7kg | Approximately 0.3kg | Approximately 0.2kg | Approximately 2.8kg | 45~85%RH (no condensation) | 0~40℃ (no freezing) | Operation method: Cutting off heater power by thermostat |
Approximately 570VA (at AC100V) Inrush current 30 A or less |
AC90~110V (50/60Hz shared) *Rated AC100V *Includes power supply voltage fluctuations |
0.025 seconds |
0.00~300.00℃ *Sensor cannot be calibrated below room temperature |
3 minutes |
± 0.1℃ (≦100℃) ± 0.2℃ (≦200℃) ± 0.4℃ (≦300℃) |
Pt100 | ±0.1℃ |
Sensor connector: AC250V (1 minute) between ground terminals Heater connector: AC1,000V (1 minute) between ground terminals |
Sensor connector: AC1,000V (1 minute) between ground terminals Heater connector: AC1,000V (1 minute) between ground terminals Power terminal: Between ground terminal AC1,500V (1 minute) |