All Categories

History

Hybrid recorder SBR-EW100 / EW180-SBR-EW101-1
Hybrid recorder SBR-EW100 / EW180-RKC INSTRUMENT

Hybrid recorder SBR-EW100 / EW180
RKC INSTRUMENT

RKC INSTRUMENT's Response Status

Response Rate

100.0%

Response Time

12.6hours

Relatively Fast Response


About This Product

Adds a sense of perfection to the ease of paper records.

■Multi-display allows you to check data at a glance

- You can choose from 80 different display patterns, including digital display of measured values ​​and bar graph display. - Up to 15 types of display patterns can be registered and the display can be switched with one touch using the "DISP" key.

■Two sizes available

Two types of front dimensions are available: 144 x 144 mm and 288 x 288 mm.

■Achieving excellent operability

- Equipped with a large VFD full dot display. - Equipped with a function to take out recording paper during recording. - Interactive settings using navigation display. ・Equipped with high-brightness LED internal lighting as standard.

■Keeps compatibility with conventional products (EY series)

・Conventional EY series spare parts (pens, cartridges, recording paper) can continue to be used as is. ・Both the back terminal arrangement (*) and instrument depth size follow the conventional EY series. (*) Includes some improvements

■Equipped with functions that meet diverse needs

・Various lineup (1, 2, 3, 4 pens, 6, 12, 18, 24 dots) *12, 18, 24 dots are available only in SBR-EW180 ・Achieved a measurement cycle of 1 second with the high-speed recording 6-dot model. ・Universal input specifications. - Compatible with special sensor types (35 types including PLⅡ, PR40-20, etc.). (optional) ・Can be equipped with calculation functions. (optional) ・Number of calculation channels: Pen model 8ch, dot model max. 12ch (EW100) max. 24ch (EW180) ・Calculation type: General-purpose calculation, statistical calculation *Calculation results can be recorded on chart paper - Compatible with general purpose communication RS-422A/485. (optional)

■You can write notes and check data even while recording.

The chart cassette is additionally equipped with a function to take out recorded chart paper. Even while recording, you can easily check past data and write notes.

■High precision and high reliability of mechanical drive parts

The pen servo section has been changed from the conventional belt drive system to an ultra-compact stepping motor and rack & pinion system. Furthermore, the position detection method has been changed to a non-contact optical encoding method. Achieved longer life and improved reproducibility. *Other product numbers are available.

  • Product

    Hybrid recorder SBR-EW100 / EW180

Share this product


540+ people viewing

Last viewed: 3 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

10 Models of Hybrid recorder SBR-EW100 / EW180

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Power frequency Brightness settings Setting screen display Recording paper length Chart paper feed speed change Chart paper feed speed Recording accuracy Recording method Recording cycle Language display Display update cycle Display method Display content Paper feed accuracy RTD input Measurement period Effective recording width Dot model resolution Rated power supply voltage Input calculation function Input resistance Input external resistance Usable power supply voltage range Operating temperature and humidity range Memory backup Filter function Burnout Bar graph display Type Pen model dead band
Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW101-1

SBR-EW101-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

16m (folding)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 100mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (101×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

100mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW100 1 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW102-1

SBR-EW102-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

16m (folding)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 100mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (101×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

100mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW100 2 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW103-1

SBR-EW103-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

16m (folding)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 100mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (101×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

100mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW100 3 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW104-1

SBR-EW104-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

16m (folding)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 100mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (101×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

100mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW100 4 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW106-1

SBR-EW106-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

16m (folding)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

1~1,500mm/h (1mm step)

Measurement accuracy ± (0.3% of recording span)
Record span: 100mm

6 color wire dots

6 points/10 seconds *Fastest recording cycle

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (101×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

1 second/6 points

100mm

0.1mm

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Moving average
ON/OFF can be specified for each channel
*Moving average number of times: Specify from 2 to 16 times

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW100 6 RBI recorder

-
Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW181-1

SBR-EW181-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

20m (foldable)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 180mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (181×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

180mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW180 1 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW182-1

SBR-EW182-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

20m (foldable)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 180mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (181×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

180mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW180 2 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW183-1

SBR-EW183-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

20m (foldable)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 180mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (181×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

180mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW180 3 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW184-1

SBR-EW184-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

20m (foldable)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

5~12,000mm/h (82 steps)

Measurement accuracy ± (0.3% of recording span)
Record span: 180mm

Disposable felt pen, plotter pen

Continuous recording for each channel

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (181×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

0.125 seconds for all channels

180mm

-

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Signal damping
ON/OFF can be specified for each channel
*Time constant: Specify from 2, 5, or 10 seconds

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW180 4 pen recorder

Less than 0.2% of recording span

Hybrid recorder SBR-EW100 / EW180-Part Number-SBR-EW186-1

SBR-EW186-1

Available upon quote

50Hz/60Hz (automatic switching)

Brightness level configurable

Interactive display of various recorder settings
*When setting various settings, navigation to support settings is displayed at the bottom of the display.

20m (foldable)

Speed ​​1/Speed ​​2 can be switched by remote control (optional)

1~1,500mm/h (1mm step)

Measurement accuracy ± (0.3% of recording span)
Record span: 180mm

6 color wire dots

6 points/10 seconds *Fastest recording cycle

Japanese display

AUTO: 1~5 seconds
MAN: 2 seconds (pen model), linked to measurement cycle (dot model)
*AUTO/MAN switchable

VFD (181×16 dot matrix)

Digital display, bar graph, flag display, DI/DO status, date/time, etc. *Display settings can be registered in up to 15 formats from approximately 80 formats.

Within ±0.1% *However, if the feed is more than 1,000 mm, the print scale is based on the recording paper.

1 wire 10Ω or less (all 3 wires must be equal)

1 second/6 points

180mm

0.1mm

100~240VAC (automatic switching)

Channel difference, linear scaling, square root calculation (with low cut function), bias (arithmetic addition)

Voltage range below 200mVDC and thermocouple: 10MΩ or above
Voltage range above 2VDC: approx. 1MΩ

DC voltage, thermocouple input: 2kΩ or less

90~132VAC, 180~264VAC

0~50℃, 20~80%RH (at 5~40℃)

Set value and clock operation protected by lithium battery
Lithium battery life: 10 years (23±2℃ 55±10%RH)

Moving average
ON/OFF can be specified for each channel
*Moving average number of times: Specify from 2 to 16 times

Detected for TC, DCV (1~5V)
*Detection ON/OFF switchable for each channel
*1~5V range burnout below 0.2V

Measured value: Can display left/right edge reference or center zero bar graph *Can be specified for each channel
Alarm: Alarm set point display and alarm point display

SBR-EW180 6 RBI recorder

-

Click on the part number for more information about each product

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Data Loggers Products

Other products of RKC INSTRUMENT

Reviews shown here are reviews of companies.


View more products of RKC INSTRUMENT

About Company Handling This Product

Response Rate

100.0%


Response Time

12.6hrs

Company Overview

RKC Instrument has been continually growing for over half a century and strives to be a solution provider of m...

See More

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree