All Categories

History

Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P-RI-2000P
Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P-Lab System Equipment Co., Ltd.

Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P
Lab System Equipment Co., Ltd.


About This Product

Lab System's original differential refractive index detector boasts industry-leading separation maintenance performance.

■Features

・A new method is used to suppress peak dispersion in general preparative separation and recycling preparative separation. - Optical zero adjustment can be easily performed from the front panel. -Uses a high-intensity LED lamp as the light source, eliminating baseline drift due to lamp heat generation. - Temperature control is applied to the optical system, allowing for fast and stable measurements. (RI-2000AP/P: approx. 30 minutes, RI-2000A: approx. 120 minutes from temperature control ON) ・The RI-2000AP has flow paths for analysis and preparative separation, so one unit can be used for both analysis and preparative separation without replacing the flow cell. -Easily switch between analysis and preparative flow paths and perform reference purge operations. ・In case of a liquid leak, a ``liquid leak detection (alarm) unit'' (optional) can be installed to prevent the accident from spreading. ・We can also manufacture "inert specification machines" that do not use any metal in the wetted parts. (Special order product)

■The biggest feature of the RI-2000 series is safe valve flow switching.

・It is possible to maintain separation of drainage liquid. The detected data and the components of the waste liquid are almost the same. ・Safe and unbreakable as almost no high pressure is generated during switching. Detection data is also stable. By making special improvements to our valves, we have solved the problem of high pressure generation and are now able to collect safe and stable data while maintaining drainage fluid separation. Reliable quality control

■Traceability system

All of the differential refractive index detectors in the laboratory system are tested by skilled technicians in accordance with the procedures recommended by the Japan Analytical Instruments Association standard "Method for expressing the performance of high-performance liquid chromatographs" (JAIMAS 0005-1984). Calibration is performed using equipment that has been regularly certified by a measuring instrument manufacturer or certified calibration company. This allows us to say that we have traceability to national standards.

■Traceability system chart and calibration certificate can also be issued.

The laboratory system's differential refractive index detector undergoes rigorous standard testing, but we are also able to issue traceability system diagrams and calibration certificates so that it can be used in fields that require extremely high precision, such as pharmaceuticals and foods. It has become.

  • Product

    Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P

Share this product


110+ people viewing

Last viewed: 5 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P

Image Part Number Price (excluding tax) Power supply Weight Recorder output Purpose Measurement method Measuring range Temperature control function Channel capacity Wetted parts material Refractive index range Dimensions (mm) Usage flow rate Linearity range Noise level Drift Cell pressure resistance Cell capacity Inte output Option
Differential refractive index detector for liquid chromatography RI-2000 series RI-2000P-Part Number-RI-2000P

RI-2000P

Available upon quote

AC85~240V (50/60Hz), 100VA

Approximately 10kg

±10mV/FS (with marker)

For bulk separation

Deflection type

0.2~100×10^4RIU 9 steps switching

Set to 33℃ (with 50℃ temperature protector)

IN~cell: approx. 640μl
Sample cell~OUT: approx. 200μl
Reference cell~OUT: approx. 260μl

Stainless steel, Teflon, quartz glass, Tefzel, PEEK (excluding stainless steel for inert specification machines)

1.00~1.75

260 (W) ×410 (D) ×160 (H) Excluding protrusions

~150ml/min (when using H2O)

600μRIU or more

40×10^8 RIU or less

2×10^6RIU/h or less (under conditions of water filling and no room temperature fluctuation)

0.49MPa

9μl

MAX±3.5V

Liquid leak detection unit

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Refractometers Products

Other products of Lab System Equipment Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Lab System Equipment Co., Ltd.

About Company Handling This Product

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree