All Categories
History
Product
High-performance radiation therapy electrometer SUPERMAXHandling Company
Chiyoda Technol Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | Nonlinearity | Charge collection trigger | Charge collection timer | Charge collection continue | Power supply | Long term stability | Suitability | Display range RATE | Display range CHARGE | Accuracy | Size | Stabilization time | Reaction time | Applied voltage | Output | Reproducibility | Internal memory | Input | Range | Real time clock | User settings | Display | Zero drift | Zero shift | Zeroing | Option |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
SUPERMAX (REF90018) |
Available upon quote |
±0.25% (IEC60731 recommended: ±1.0%) |
Automatic measurement start/stop/reset function, data storage based on user-set thresholds. (Start: 0.2-9.9pA, Stop 0.1-9.8pA) |
Measurement time setting by user (range: 1-600 seconds: 1 second interval setting) |
Continuous manual measurement |
Input 100~240VAC, 50/60Hz, max. 0.5A, with power adapter |
±0.5% (more than 1 year) |
CE93/42/EEC Reference class according to IEC60731 |
Low range 0.001pA-500.0pA (resolution 1fA), High range 0.001nA-500.0nA (resolution 1pA) |
Low range 0.001pC-999.9μC (resolution 1fC), High range 0.001nC-999.9μC (resolution 1pC) |
±0.3V |
Length x width x height: 21.1cm x 26.7cm x 8.1cm, weight: 2.5kg |
±0.1% (IEC recommendation: ±0.5%/1 hour (based on measurements taken at 15 minutes and 6 hours)) |
High range: 3 seconds or less (IEC60731 recommended: 3 seconds or less), Low range: 12 seconds or less |
Standard ±1,000V |
USB port x 2 |
±0.1% (IEC recommendation: ±0.5%) |
Number of saved data <100 data, number of chamber library registrations <100 |
BNC-2 lug/Triaxial connector x 2 |
User selection - High range (nC) or Low range (pC) |
Internal clock to manage measurement time |
-1,000~-100, 0, 100~1,000V (1V step) |
6.4 inch color TFT touch screen |
Minimum effective reading scale of ±0.25% (low: ±0.001pA or less/high: ±0.001nA or less) (IEC60731 recommended: ±0.5%) |
Minimum effective reading scale of ±0.25% (low: ±0.001pA or less/high: ±0.001nA or less) (IEC60731 recommended: ±0.5%) |
Auto zero function |
It is also possible to change to TNC connector. |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
60.3hrs
Company Review
5.0