Radioactivity measurement device using a well-shaped ionization chamber Digital CurimeterHandling Company
Chiyoda Technol Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | High pressure display | Power supply | Mass | Display range | Measurement result display | Measurement accuracy | Measuring range | Measuring nuclide | Measurement time remaining amount display | Measurement time | Unit of measurement | Detector well volume | Detection method | Dosed liquid volume calculation | External communication output | Half-life calculation | Half-life graph | Reproducibility | Usage environment | Background subtraction | Weight time |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
IGC-8B |
Available upon quote |
Can be |
Main unit: AC100V, 50/60Hz, max. 40VA, Dot printer (optional): AC100V, 50/60Hz, approx. 7VA, Label printer (optional): AC100V, 50/60Hz, approx. 45VA |
Detection part: Approx. 21kg, Measuring part: Approx. 4kg, Dot printer (optional): Approx. 0.5kg, Label printer (optional): Approx. 1.6kg |
0.000kBq~999.9GBq (0.00μCi~999.9Ci) Automatic range switching |
Nuclide, radioactivity amount, half-life graph, half-life calculation |
Measurement time automatic mode: ±5% (99mTc approximately 19MBq or more), measurement time specified mode (with cumulative average): ±5% (measurement time 2 seconds, number of repeats 20 times, positron nuclide, approximately 4.2MBq or more) |
Measurement time automatic mode: 0.01MBq to approx. 100GBq (0.1μCi to approx. 3Ci), measurement time specified mode (with cumulative average): 0.01MBq to approx. 22GBq (0.1μCi to approx. 0.6Ci) |
Γ (X) ray emitting nuclides of 30keV or higher, 21 registered nuclides with conversion constants set, 5 registered nuclides with no conversion constants set. |
Can be |
Measurement time automatic mode (at 99mTc): 370MBq or more approximately 2~5 seconds, 11~370MBq approximately 10 seconds, 11MBq or less 120 seconds or less, measurement time specification mode: 2~100 seconds (can be set in 1 second increments) |
Bq, Ci |
Approximately φ44mm×270mm |
Well-type ionization chamber filled with argon gas |
Can be |
Can be |
Can be |
Half-life calculation results can be graphed |
Measurement time automatic mode: ±1% (99mTc approximately 19MBq or more), measurement time specified mode (with integrated average): ±1% (measurement time 2 seconds, number of repeats 20 times, positron nuclide, approximately 4.2MBq or more) |
Temperature: 15~35℃, Humidity: 30~75% (no condensation), Air pressure: 75~106kPa |
Can be |
Can be set from 0 to 999 minutes when setting repeat |