Glow discharge emission spectrometer GDS900 typeHandling Company
LECO Japan LLCCategories
Image | Part Number | Price (excluding tax) | Sound pressure level | Power capacity | Sample cooling | Installation conditions | Vacuum pump | Light source | Amount of heat | Detector | Dimensions/weight | Resolution | Optical department | Gas used Light emitting part/spectroscope | Gas used Purge gas | Gas used Driving gas | Purge system | Spectral range |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
GDS900 |
Available upon quote |
53 dBa (according to IEC / EN 61010-1, maximum value at operator position) |
Single phase 200 V (±10%, boosted to 230 V), 50/60 Hz, 14 A (max.) |
Internal closed-loop circulation cooling system |
Operating temperature: 18℃~30℃ |
External vacuum pump (optional: soundproof cover) |
4 mm diameter grim type glow discharge source (optional: 2 mm diameter and 7 mm diameter) |
1,800 BTU/hr |
16,000 active pixels (14.0 mm W x 0.9 mm H pixel size) CCD detector |
Width: 53.3 cm x Depth: 215.9 cm (equipment only) x Height: 129.5 cm / 317.5 kg |
50pm (0.050nm) |
Focal length: 0.75 m, 900 lines/mm, Littrow mounting |
Argon (99.998%), 0.28 MPa ±10% |
Nitrogen (99.99%), 0.14 MPa ±10% |
Air or nitrogen (no moisture or oil), 0.41 MPa ±10% |
0.25 L/min high purity nitrogen (optional: recirculation system) |
160nm~460nm |