All Categories
History
Response Rate
100.0%
Response Time
39.2hours
Product
Low-temperature constant temperature water circulation device Unithermo Ace Eco NCC-3000/3100 seriesHandling Company
Tokyo Rikakikai Co., Ltd.Categories
Click on the part number for more information about each product
Image | Part Number | Price (excluding tax) | Power supply voltage | Power input | Mass | Temperature control accuracy | Temperature setting range | Temperature setting/display | Temperature control method | Temperature sensor | Water tank capacity | Compatible circulating water | Safety features | External circulation capacity (50Hz) Maximum flow rate | External circulation capacity (50Hz) Maximum head | External circulation nozzle | External input/output | External dimensions (mm) | Cooling capacity (at liquid temperature) -10℃ | Cooling capacity (at liquid temperature) 30℃ | Cooling capacity (at liquid temperature) 10℃ | Cooling/circulation method | Freezer/refrigerant | Supply water | Operating ambient temperature range | Attached functions | Heater |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
NCC-3000A |
Available upon quote |
AC200V single phase 50/60Hz |
13.5A, 2.7kVA |
Approximately 53kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Water supply |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
14.7L/min |
10.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
660W (567kcal/h) |
1,600W (1,376kcal/h) |
1,230W (1,057kcal/h) |
Air-cooled/closed system directional circulation |
Air-cooled output 550W/R410C |
- |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
![]() |
NCC-3000B |
Available upon quote |
AC200V single phase 50/60Hz |
14A, 2.8kVA |
Approximately 53kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Water supply |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
21.8L/min |
19.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
500W (430kcal/h) |
1,440W (1,238kcal/h) |
1,070W (920kcal/h) |
Air-cooled/closed system directional circulation |
Air-cooled output 550W/R410C |
- |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
NCC-3000C |
Available upon quote |
AC200V single phase 50/60Hz |
14A, 2.8kVA |
Approximately 55kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Clean water and pure water |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
14.7L/min |
10.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
685W (589kcal/h) |
1,840W (1,582kcal/h) |
1,400W (1,204kcal/h) |
Air-cooled/closed system directional circulation |
Air-cooled output 550W/R410C |
- |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
|
NCC-3000D |
Available upon quote |
AC200V single phase 50/60Hz |
14.5A, 2.9kVA |
Approximately 55kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Clean water and pure water |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
21.8L/min |
19.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
600W (516kcal/h) |
1,680W (1,444kcal/h) |
1,240W (1,061kcal/h) |
Air-cooled/closed system directional circulation |
Air-cooled output 550W/R410C |
- |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
|
![]() |
NCC-3100A |
Available upon quote |
AC200V single phase 50/60Hz |
13.5A, 2.7kVA |
Approximately 49kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Water supply |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
14.7L/min |
10.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
730W (627kcal/h) |
2,850W (2,451kcal/h) |
1,750W (1,505kcal/h) |
Water cooling/closed system directional circulation |
Water-cooled output 550W/R410C |
35℃ or less, 30L/min or more (supply water inlet/outlet Rc1/2) |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
NCC-3100B |
Available upon quote |
AC200V single phase 50/60Hz |
14A, 2.8kVA |
Approximately 49kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Water supply |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
21.8L/min |
19.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
620W (533kcal/h) |
2,700W (2,322kcal/h) |
1,650W (1,419kcal/h) |
Water cooling/closed system directional circulation |
Water-cooled output 550W/R410C |
35℃ or less, 30L/min or more (supply water inlet/outlet Rc1/2) |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
|
NCC-3100C |
Available upon quote |
AC200V single phase 50/60Hz |
13.5A, 2.7kVA |
Approximately 52kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Clean water and pure water |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
14.7L/min |
10.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
820W (705kcal/h) |
2,850W (2,451kcal/h) |
1,750W (1,505kcal/h) |
Water cooling/closed system directional circulation |
Water-cooled output 550W/R410C |
35℃ or less, 30L/min or more (supply water inlet/outlet Rc1/2) |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
|
NCC-3100D |
Available upon quote |
AC200V single phase 50/60Hz |
14A, 2.8kVA |
Approximately 52kg |
±0.1~0.2℃ |
−10~80℃ |
Sheet key input/digital display, minimum digit 0.1℃ |
Inverter frequency control |
Platinum resistance thermometer Pt100Ω |
3.6L (liquid volume 2.8L) |
Clean water and pure water |
Temperature controller self-diagnosis function, refrigerator overload, sensor abnormality, refrigerator high pressure switch, refrigerator protection timer, float switch, SSR short circuit, heater disconnection, circulation pump thermal protector, antifreeze flow rate switch, etc. |
21.8L/min |
19.4m |
Both discharge and return ports (Rc3/8 female thread) outer diameter 14mm hose port |
Alarm output terminal: No-voltage contact (A contact) output 250V/DC30VMax.3A |
340W×440 (452) D×852H |
570W (490kcal/h) |
2,690W (2,313kcal/h) |
1,720W (1,479kcal/h) |
Water cooling/closed system directional circulation |
Water-cooled output 550W/R410C |
35℃ or less, 30L/min or more (supply water inlet/outlet Rc1/2) |
5~35℃ |
Emergency stop button, variable independent overheating preventer |
2kW (stainless steel) |
Click on the part number for more information about each product
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
39.2hrs