All Categories
History
Response Rate
100.0%
Response Time
41.7hours
Product
Cooling water circulation system Cool Ace CA-2610/2610S typeHandling Company
Tokyo Rikakikai Co., Ltd.Categories
Click on the part number for more information about each product
Image | Part Number | Price (excluding tax) | Power input/power voltage | Power utility | Mass | Emergency stop button | Temperature control accuracy | Temperature setting range | Temperature setting/display | Temperature control | Temperature sensor | Aquarium material | Water tank dimensions (mm) | Water tank capacity | Circulation pump | Safety features | External circulation ability (50/60Hz) | External circulation nozzle | External input/output | External dimensions (mm) | Cooling capacity (at liquid temperature) 20℃ | Cooling capacity (at liquid temperature) 10℃ | Cooling/circulation method | Cooling coil | Freezer/refrigerant | Operating ambient temperature range | Attached functions | Type |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CA-2610 |
Available upon quote |
10A, 2.0kVA・AC200V single phase 50/60Hz |
AC200V single phase 15A capacity |
Approximately 84kg |
- |
±2℃ or less |
−10~35℃ (without heater) |
Sheet key input/digital display |
Capillary tube capacity switching control + refrigerator ON-OFF control |
Thermistor |
Stainless |
Φ280W×270H |
16. 5L (liquid volume 14L) |
Magnet pump output 41/59W (50/60Hz) |
Earth leak/overcurrent breaker, circulation pump thermal protector, temperature rise lower limit alarm, temperature controller self-diagnosis function, sensor abnormality, refrigerator overload, refrigerator high pressure switch, refrigerator protection timer, power outage alarm |
Maximum flow rate 23.5/27L/min, maximum head 9.5/13m (Blowout pressure: 0.09/0.13MPa) |
Both discharge and return ports (Rc1/2 female thread) outer diameter 16mm hose port |
External temperature sensor terminal, temperature recorder output terminal, alarm output terminal Alarm items: Temperature rise lower limit, temperature controller self-diagnosis function, sensor abnormality, refrigerator overload, refrigerator high pressure switch, power outage alarm |
450 (555) W×515D×1,010 (1,025) H |
3,000W (2,580kcal/h) |
2,500W (2,150kcal/h) |
Air-cooled/closed system directional circulation |
Stainless |
Air-cooled output 1.1kW/R407C |
5~35℃ |
Bypass valve, flow control valve |
- | |
![]() |
CA-2610S |
Available upon quote |
10A, 2.0kVA・AC200V single phase 50/60Hz |
AC200V single phase 15A capacity |
Approximately 84kg |
Standard equipment |
±2℃ or less |
−10~35℃ (without heater) |
Sheet key input/digital display |
Capillary tube capacity switching control + refrigerator ON-OFF control |
Thermistor |
Stainless |
Φ280W×270H |
16. 5L (liquid volume 14L) |
Magnet pump output 41/59W (50/60Hz) |
Earth leak/overcurrent breaker, circulation pump thermal protector, temperature rise lower limit alarm, temperature controller self-diagnosis function, sensor abnormality, refrigerator overload, refrigerator high pressure switch, refrigerator protection timer |
Maximum flow rate 23.5/27L/min, maximum head 9.5/13m (Blowout pressure: 0.09/0.13MPa) |
Both discharge and return ports (Rc1/2 female thread) outer diameter 16mm hose port |
External temperature sensor terminal, temperature recorder output terminal, alarm output terminal Alarm items: lower limit of temperature rise, temperature controller self-diagnosis function, sensor abnormality, refrigerator overload, refrigerator high pressure switch |
450 (555) W×515D×1,010 (1,025) H |
3,000W (2,580kcal/h) |
2,500W (2,150kcal/h) |
Air-cooled/closed system directional circulation |
Stainless |
Air-cooled output 1.1kW/R407C |
5~35℃ |
Bypass valve, flow control valve |
With emergency stop button |
Click on the part number for more information about each product
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
41.7hrs