All Categories

History

UHPC series precision charge/discharge evaluation system-UHPC-2A
UHPC series precision charge/discharge evaluation system-UHPC-10A
UHPC series precision charge/discharge evaluation system-UHPC-20A
UHPC series precision charge/discharge evaluation system-Toyo Technica Co., Ltd.

UHPC series precision charge/discharge evaluation system
Toyo Technica Co., Ltd.

Toyo Technica Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

5.0hours

Very Fast Response


About This Product

■Summary

The UHPC series is an abbreviation for Ultra High Precision Coulometry, which is a measurement technology for evaluating the lifespan of lithium-ion batteries devised by Prof. Jeff Dahn of Dalhousie University in Canada. This system performs precision charge/discharge tests with a stability of 10 ppm or less and a measurement accuracy of 50 ppm or less, making it possible to evaluate side reactions and life expectancy of lithium-ion batteries due to slight differences in the amount of coulombs at the initial stage. Masu.

■Features

High precision coulometry (HPC) is an evaluation method first proposed by Professor Jeff Dahn of Dalhalsey University in Canada in 2010, which accurately measures the coulombic efficiency (CE) of a battery and uses it to quantify side reactions. We perform evaluation and analysis of deterioration factors. By using this method, it is possible to identify slight differences in Coulombic efficiency at the initial stage of charge/discharge measurements, making it possible to evaluate battery performance and predict battery life in a short time.

■Highly accurate coulombic efficiency measurement

The founding members of NOVONIX have been involved in Prof. Dahn's HPC measurement system development project, so they have a deep understanding of equipment overview, advantages of evaluation methods, and data analysis, and have successfully commercialized the HPC measurement system. NOVONIX's UHPC series precision charge/discharge evaluation system has a system that maintains the cell and measuring device at a constant temperature, and by minimizing voltage drift due to temperature changes, it has high stability with a typical value of 10 ppm or less, and 50 ppm or less. It is possible to perform charge/discharge measurements with high accuracy and calculate coulombic efficiency with high accuracy. High-precision coulometry is difficult to measure with general charging/discharging equipment due to various error factors, and the main error factors are listed below.

■Others

・Current control accuracy ・Voltage measurement accuracy ・Cell temperature *As a general rule, keep the C rate below C/10 to suppress heat generation. ・Measurement error due to temperature unevenness in the measuring instrument *Measurement made possible by the development of ultra-high precision coulometry

  • Product

    UHPC series precision charge/discharge evaluation system

Share this product


230+ people viewing

Last viewed: 3 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

3 Models of UHPC series precision charge/discharge evaluation system

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Current measurement Measurement accuracy Current measurement Measurement resolution Current resolution Current output control accuracy Current range Voltage measurement Measurement accuracy Voltage measurement Measurement resolution Voltage resolution Voltage output control accuracy Voltage range Temperature coefficient Resolution Noise Number of channels Sampling rate
UHPC series precision charge/discharge evaluation system-Part Number-UHPC-2A

UHPC-2A

Available upon quote

0.005% of range
0.002% of range (at 23±5℃)

20ppm or less in the microwave

0.002% or less of range

0.01% of range

200µ~2A (5 ranges)

100µV (<250mV)
200µV (>250mV)

<50nV (<250mV)
<1µV (>250mV)

1µV

100µV (<250mV)
200µV (>250mV)

250mV,5V

0.002% of range (at 23±5℃)

0.01℃

0.002% of range

8

6Hz

UHPC series precision charge/discharge evaluation system-Part Number-UHPC-10A

UHPC-10A

Available upon quote

0.005% of range
0.002% of range (at 23±5℃)

20ppm or less in the microwave

0.002% or less of range

0.01% of range

100µA~10A (6 ranges)

200µV

1µV

100µV or less

200µV

0~5V

0.002% of range (at 23±5℃)

0.01℃

0.002% of range

8

6Hz

UHPC series precision charge/discharge evaluation system-Part Number-UHPC-20A

UHPC-20A

Available upon quote

0.005% of range
0.002% of range (at 23±5℃)

20ppm or less in the microwave

0.002% or less of range

0.01% of range

2A~20A (2 ranges)

200µV

1µV

50µV or less

200µV

0~5V

0.002% of range (at 23±5℃)

0.01℃

0.002% of range

4

6Hz

Click on the part number for more information about each product

Other products of Toyo Technica Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Toyo Technica Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

5.0hrs

  • Japan