About This Product

IN5+ series is a digital radiation thermometer with fast response (fastest 10msec). With a detection wavelength of 8 to 14 μm, it is possible to measure the temperature of a wide variety of materials other than non-metallic surfaces.

■Applications

・Digital radiation thermometer for low temperature ・Food, rubber, plastic, paper, printing, ceramics, paint, asphalt, textiles, wood, non-metallic objects - Suitable for measuring non-metallic surfaces.

■Features

・High-speed response (fastest: 10msec) is possible, making it suitable for measuring the temperature of samples that experience rapid temperature changes or samples that move at high speed. ・Since small spot measurement is achieved, even small samples can be measured. IN5-L+ realizes a small spot of φ1/100mm. ・High accuracy and stability with fully digital circuit, scaling freely over temperature range. ・Position can be easily confirmed with laser pilot light. ・Since it has a built-in maximum value holding circuit, it is suitable for measuring moving objects. ・Data can be recorded and parameters can be set using the standard measurement software (for Windows).

■Specifications

・Can be replaced with previous models with scaling function of minimum 50℃ width. - Analog output can be selected from current 0/4 to 20mA, voltage 0 to 5V, and thermocouple J/K compatible output. ・The interface is RS232 or RS485 with bus connection. (specified at time of order) ・Compact design of φ50×115mm (Main body material is SUS)

  • Product

    Digital radiation thermometer IN5+ IN5-H+ IN5-L+

Share this product


1840+ people viewing

Last viewed: 13 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

3 Models of Digital radiation thermometer IN5+ IN5-H+ IN5-L+

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Weight Communication function Measurement software Insulation measures Temperature range Power consumption Detection wavelength Detector Emissivity correction e Response time t90=minimum value External dimensions Ambient temperature Installation direction Resolution Output signal Reproducibility Optical system Protection class Storage temperature Power supply Location confirmation Lens Housing Type Scaling Warm up time CE mark
Digital radiation thermometer IN5+ IN5-H+ IN5-L+-Part Number-IN5+

IN5+

Available upon quote

410g

RS232C or RS485 (addressable), half duplex, baud rate 2,400 ~ 19,200

InfraWin attached (works on Windows PC)

Power supply circuit, interface, and analog output signals are isolated from each other and from the main unit.

-32~900℃

Max. 70mA

8~14μm

Thermopile

0.20~1.00 (can be changed in 0.01 steps)

80msec
Body: 0.5s, 1s, 2s, 5s
Via communication interface: 0.5s, 1s, 2s, 5s, 10s, 30s

Φ49.5×114.5mm, body material is SUS

0 ~ 63℃, but no condensation
The laser pilot light stops when the internal temperature of the radiation thermometer is 50℃ or higher.

Omnidirectional

0.1℃

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less (at 24V DC)

0.3% of measured value or 0.6℃, whichever is greater

Fixed focus, optics=100, 300, 800 to choose from

IP65 (DIN 40 050)

-20~+70℃

DC18~30V, voltage ripple <0.5V

Laser pilot light IEC608251-3-4
(IEC608251-3-4: Laser class 2, maximum output <1mW, 650nm)

ZnS

Stainless steel

80msec: Standard type

Any setting possible within the temperature range, minimum scaling range 51℃

15 minutes or more

Compliant with EU Directive on Electromagnetic Immunity

Digital radiation thermometer IN5+ IN5-H+ IN5-L+-Part Number-IN5-H+

IN5-H+

Available upon quote

410g

RS232C or RS485 (addressable), half duplex, baud rate 2,400 ~ 19,200

InfraWin attached (works on Windows PC)

Power supply circuit, interface, and analog output signals are isolated from each other and from the main unit.

-32~900℃

Max. 70mA

8~14μm

Thermopile

0.20~1.00 (can be changed in 0.01 steps)

10msec
Body: 0.5s, 1s, 2s, 5s
Via communication interface: 0.5s, 1s, 2s, 5s, 10s, 30s

Φ49.5×114.5mm, body material is SUS

0 ~ 63℃, but no condensation
The laser pilot light stops when the internal temperature of the radiation thermometer is 50℃ or higher.

Omnidirectional

0.1℃

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less (at 24V DC)

0.3% of measured value or 0.6℃, whichever is greater

Fixed focus, optics=100, 300, 800 to choose from

IP65 (DIN 40 050)

-20~+70℃

DC18~30V, voltage ripple <0.5V

Laser pilot light IEC608251-3-4
(IEC608251-3-4: Laser class 2, maximum output <1mW, 650nm)

ZnS

Stainless steel

10msec: High speed type

Any setting possible within the temperature range, minimum scaling range 51℃

15 minutes or more

Compliant with EU Directive on Electromagnetic Immunity

Digital radiation thermometer IN5+ IN5-H+ IN5-L+-Part Number-IN5-L+

IN5-L+

Available upon quote

410g

RS232C or RS485 (addressable), half duplex, baud rate 2,400 ~ 19,200

InfraWin attached (works on Windows PC)

Power supply circuit, interface, and analog output signals are isolated from each other and from the main unit.

0~900℃

Max. 70mA

8~14μm

Thermopile

0.20~1.00 (can be changed in 0.01 steps)

180msec
Body: 0.5s, 1s, 2s, 5s
Via communication interface: 0.5s, 1s, 2s, 5s, 10s, 30s

Φ49.5×114.5mm, body material is SUS

0 ~ 63℃, but no condensation
The laser pilot light stops when the internal temperature of the radiation thermometer is 50℃ or higher.

Omnidirectional

0.1℃

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less (at 24V DC)

0.3% of measured value or 0.6℃, whichever is greater

Fixed focus, optics=100, 300, 800 to choose from

IP65 (DIN 40 050)

-20~+70℃

DC18~30V, voltage ripple <0.5V

Laser pilot light IEC608251-3-4
(IEC608251-3-4: Laser class 2, maximum output <1mW, 650nm)

ZnS

Stainless steel

180msec: small spot type

Any setting possible within the temperature range, minimum scaling range 51℃

30 minutes or more

Compliant with EU Directive on Electromagnetic Immunity

Click on the part number for more information about each product

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Radiation Thermometers Products

Other products of Yamazato Sangyo Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Yamazato Sangyo Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

275.2hrs

  • Japan