All Categories

History

Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-IS6
Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-Yamazato Sangyo Co., Ltd.

Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV
Yamazato Sangyo Co., Ltd.

Yamazato Sangyo Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

275.2hours


About This Product

IS6/IGA6 IS6-TV/IGA6-TV is a fully digital radiation thermometer. Adopts a high-speed response of 120 μsec and a movable focus lens. The temperature of metals, ceramics, etc. can be measured with high precision.

■Applications

・Metal processing ・Semiconductor production process ・Glass mold ·ceramic ・Molted metal ・Forging

■Features

・High-speed response: 120μsec ・Small spot: from φ0.6mm/210mm ・With 4-digit LED display ・Adopts a movable focus lens ・High accuracy: 0.3% of measured value + 2℃ (<1,500℃) ・Wide range: Minimum 50°C width scaling function makes it easy to replace conventional models ・Position confirmation: Please select from 3 methods. (Remote monitoring with laser pilot light, direct view finder, and video output type) ・DC4 to 20mA output without a separate converter ・RS485 interface allows long-distance communication with PCs and PLCs ・Built-in maximum value holding circuit: For measuring moving objects ・Compact design of φ50×127.5mm, body material is SUS ・Data recording and parameter setting using the standard included dedicated measurement software

■Application

・Temperature measurement of low emissivity objects ・Measuring the temperature inside the vacuum furnace through the window glass ・Temperature measurement of high-speed moving objects ・Molten metal surface temperature measurement ・Temperature measurement of minute objects

  • Product

    Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV

Share this product


270+ people viewing

Last viewed: 14 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

4 Models of Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Weight Communication function Transmittance correction τ Measurement software Insulation measures Accuracy (uncertainty) Focal length Temperature range Power consumption Detection wavelength Detector Main unit LED display Emissivity correction ε Response time t90 External dimensions Ambient temperature Installation direction Resolution Output signal Reproducibility Protection class Storage temperature Power supply Location confirmation Lens focus Parameters Housing Scaling CE mark
Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-Part Number-IS6

IS6

Available upon quote

Approximately 600g

RS485 (addressable), half duplex, baud rate 1,200 ~ 115. 2k

0.050~1.000 0.001 unit

InfraWin attached (works on Windows PC)

Power circuit, interface, and analog output signals are isolated from each other

0.3% of measured value + 2℃ < 1,500℃
0.6% of measured value > 1,500℃

210mm ~ 5,000mm

MB 18: 600 ~1,800℃
MB30: 600 ~3,000℃

Max. 3W (including laser pilot light)

0.7~1.1μm

Si photodiode

4-digit LED (character height 5mm); measurement temperature, focal length

0.050~1.000 0.001 unit

120 μs (Response time automatically extended to 1ms at temperatures below 300℃ for IGA6)
Settings: 1ms, 3ms, 5ms, 10ms, 50ms, 250ms 1s, 3s, 10s

Φ49.5×127.5~185.1mm

0~70℃

Omnidirectional

0.1℃ (via communication) 1℃ (main unit LED)

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less

0.15% of measured value + 1℃

IP65 (IEC 60529)

-20~80℃

DC24V ±25% ripple less than 50mV

Choose from laser pilot light or direct view finder

Movable focus method (Please refer to the section on measurement distance and area.)

Settings via communication
Emissivity correction ε
Transmittance correction τ
response time t90
0~20mA/4~20mA switching
scaling
Maximum value holding time tCL,
°C / °F
address
baud rate

Stainless steel

Any setting possible within the temperature range, minimum scaling range 50℃

Compliant with EU Directive on Electromagnetic Immunity

Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-Part Number-IGA6

IGA6

Available upon quote

Approximately 600g

RS485 (addressable), half duplex, baud rate 1,200 ~ 115. 2k

0.050~1.000 0.001 unit

InfraWin attached (works on Windows PC)

Power circuit, interface, and analog output signals are isolated from each other

0.3% of measured value + 2℃ < 1,500℃
0.6% of measured value > 1,500℃

210mm ~ 5,000mm

MB 18: 250 ~1,800℃
MB25: 250 ~2,500℃

Max. 3W (including laser pilot light)

1.45~1.8μm

InGaAs photodiode

4-digit LED (character height 5mm); measurement temperature, focal length

0.050~1.000 0.001 unit

120 μs (Response time automatically extended to 1ms at temperatures below 300℃ for IGA6)
Settings: 1ms, 3ms, 5ms, 10ms, 50ms, 250ms 1s, 3s, 10s

Φ49.5×127.5~185.1mm

0~70℃

Omnidirectional

0.1℃ (via communication) 1℃ (main unit LED)

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less

0.15% of measured value + 1℃

IP65 (IEC 60529)

-20~80℃

DC24V ±25% ripple less than 50mV

Choose from laser pilot light or direct view finder

Movable focus method (Please refer to the section on measurement distance and area.)

Settings via communication
Emissivity correction ε
Transmittance correction τ
response time t90
0~20mA/4~20mA switching
scaling
Maximum value holding time tCL,
°C / °F
address
baud rate

Stainless steel

Any setting possible within the temperature range, minimum scaling range 50℃

Compliant with EU Directive on Electromagnetic Immunity

Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-Part Number-IS6-TV

IS6-TV

Available upon quote

Approximately 600g

RS485 (addressable), half duplex, baud rate 1,200 ~ 115. 2k

0.050~1.000 0.001 unit

InfraWin attached (works on Windows PC)

Power circuit, interface, and analog output signals are isolated from each other

0.3% of measured value + 2℃ < 1,500℃
0.6% of measured value > 1,500℃

210mm ~ 5,000mm

MB 18: 600 ~1,800℃
MB30: 600 ~3,000℃

Max. 3W (including laser pilot light)

0.7~1.1μm

Si photodiode

4-digit LED (character height 5mm); measurement temperature, focal length

0.050~1.000 0.001 unit

120 μs (Response time automatically extended to 1ms at temperatures below 300℃ for IGA6)
Settings: 1ms, 3ms, 5ms, 10ms, 50ms, 250ms 1s, 3s, 10s

Φ49.5×127.5~185.1mm

0~70℃

Omnidirectional

0.1℃ (via communication) 1℃ (main unit LED)

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less

0.15% of measured value + 1℃

IP65 (IEC 60529)

-20~80℃

DC24V ±25% ripple less than 50mV

Choose from laser pilot light or direct view finder

Movable focus method (Please refer to the section on measurement distance and area.)

Settings via communication
Emissivity correction ε
Transmittance correction τ
response time t90
0~20mA/4~20mA switching
scaling
Maximum value holding time tCL,
°C / °F
address
baud rate

Stainless steel

Any setting possible within the temperature range, minimum scaling range 50℃

Compliant with EU Directive on Electromagnetic Immunity

Digital radiation thermometer IS6/IGA6 IS6-TV/IGA6-TV-Part Number-IGA6-TV

IGA6-TV

Available upon quote

Approximately 600g

RS485 (addressable), half duplex, baud rate 1,200 ~ 115. 2k

0.050~1.000 0.001 unit

InfraWin attached (works on Windows PC)

Power circuit, interface, and analog output signals are isolated from each other

0.3% of measured value + 2℃ < 1,500℃
0.6% of measured value > 1,500℃

210mm ~ 5,000mm

MB 18: 250 ~1,800℃
MB25: 250 ~2,500℃

Max. 3W (including laser pilot light)

1.45~1.8μm

InGaAs photodiode

4-digit LED (character height 5mm); measurement temperature, focal length

0.050~1.000 0.001 unit

120 μs (Response time automatically extended to 1ms at temperatures below 300℃ for IGA6)
Settings: 1ms, 3ms, 5ms, 10ms, 50ms, 250ms 1s, 3s, 10s

Φ49.5×127.5~185.1mm

0~70℃

Omnidirectional

0.1℃ (via communication) 1℃ (main unit LED)

DC4 ~ 20mA/DC0 ~ 20mA linear output (selectable), load resistance 500Ω or less

0.15% of measured value + 1℃

IP65 (IEC 60529)

-20~80℃

DC24V ±25% ripple less than 50mV

Choose from laser pilot light or direct view finder

Movable focus method (Please refer to the section on measurement distance and area.)

Settings via communication
Emissivity correction ε
Transmittance correction τ
response time t90
0~20mA/4~20mA switching
scaling
Maximum value holding time tCL,
°C / °F
address
baud rate

Stainless steel

Any setting possible within the temperature range, minimum scaling range 50℃

Compliant with EU Directive on Electromagnetic Immunity

Click on the part number for more information about each product

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Radiation Thermometers Products

Other products of Yamazato Sangyo Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Yamazato Sangyo Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

275.2hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree