Sample chamber LMT-S series (for laser coordinate measuring machine)-C-705 (for laser coordinate measuring machine)
Sample chamber LMT-S series (for laser coordinate measuring machine)-Colette Industries Co., Ltd.

Sample chamber LMT-S series (for laser coordinate measuring machine)
Colette Industries Co., Ltd.

Colette Industries Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

116.6hours


About This Product

■Summary

This chamber is intended for non-contact measurement of thermal deformation due to temperature changes of -120℃ to 350℃ in semiconductor packages, bare chips, printed circuit boards, and various materials using a laser coordinate measuring machine.

■Accurate temperature control

Highly accurate temperature control and temperature holding accuracy of within ±0.2°C are achieved by controlling the flow rate of liquid nitrogen and the heater output of a high-precision temperature controller.

■Uniform surface temperature

The sample stage is made of highly thermally conductive pure silver that minimizes thermal distortion due to cryo treatment and annealing treatment, making it possible to maintain a uniform surface temperature.

■Uniform flatness

After applying a temperature change to the pure silver sample stage and assembling it in the chamber, we use tertiary processing to ensure flatness with mechanical precision.

■Wide adaptability

It can be attached and detached regardless of the manufacturer of various laser coordinate measuring machines.

■High quality and low price

Due to our in-house production process and thorough quality control, we are able to provide high quality products at low prices.

■Production of custom-made products

We handle the entire process in-house, starting with the initial meeting, designing, manufacturing, assembling, and testing, and we also manufacture many custom-made products.

■Details

・The water-cooled chamber case is kept at room temperature even when used at high temperatures, taking precautions against risks such as burns. - When used at low temperatures, piping through which liquid nitrogen flows can be connected with SUS taper screws to prevent cracking, falling off, or leakage, and consideration has been given to the risk of frostbite. ・Even when supplying liquid nitrogen using the N2 overpressure method, the siphon is equipped with a precision pressure reducing valve that releases the pressure when it exceeds a certain pressure, making it possible to supply liquid nitrogen safely. ・A highly safe (SUS + silicone) dedicated braided hose is used for the liquid nitrogen supply hose.

  • Product

    Sample chamber LMT-S series (for laser coordinate measuring machine)

Share this product


410+ people viewing

Last viewed: 2 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

4 Models of Sample chamber LMT-S series (for laser coordinate measuring machine)

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Weight (kg) Sample atmosphere Sample setting Sample size (diameter) Sample size (thickness) Observation window (quartz glass) Observation window (material) Observation window (thickness) Water supply capacity Measurement temperature Water cooling mechanism Main body/top cover material Dimensions (mm) Working distance Chamber model Stage material
Sample chamber LMT-S series (for laser coordinate measuring machine)-Part Number-C-705 (for laser coordinate measuring machine)

C-705 (for laser coordinate measuring machine)

Available upon quote

3.8

Atmosphere, inert gas atmosphere

Installed from the top

50mm

10mm (MAX)

50mm

Quartz glass

1.0mm

City water 2L/min

-120℃⇔350℃

Air-cooled water-cooled circulation system (OP)

Aluminum alloy

W160XD160XH53

15mm or more

S-5-50L

Silver/stainless steel

Sample chamber LMT-S series (for laser coordinate measuring machine)-Part Number-C-707 (for laser coordinate measuring machine)

C-707 (for laser coordinate measuring machine)

Available upon quote

4.2

Atmosphere, inert gas atmosphere

Installed from the top

70mm

10mm (MAX)

70mm

Quartz glass

1.0mm

City water 2L/min

-100℃⇔350℃

Air-cooled water-cooled circulation system (OP)

Aluminum alloy

W160XD160XH53

15mm or more

S-5-70L

Silver/stainless steel

Sample chamber LMT-S series (for laser coordinate measuring machine)-Part Number-C-710 (for laser coordinate measuring machine)

C-710 (for laser coordinate measuring machine)

Available upon quote

4.8

Atmosphere, inert gas atmosphere

Installed from the top

100mm

10mm (MAX)

100mm

Quartz glass

1.0mm

City water 2L/min

-80℃⇔350℃

Air-cooled water-cooled circulation system (OP)

Aluminum alloy

W190XD190XH53

15mm or more

S-5-100L

Silver/stainless steel

Sample chamber LMT-S series (for laser coordinate measuring machine)-Part Number-C-715 (for laser coordinate measuring machine)

C-715 (for laser coordinate measuring machine)

Available upon quote

6.5

Atmosphere, inert gas atmosphere

Installed from the top

150mm

10mm (MAX)

150mm

Quartz glass

1.0mm

City water 2L/min

-50℃⇔350℃

Air-cooled water-cooled circulation system (OP)

Aluminum alloy

W250XD250XH53

15mm or more

S-5-150L

Silver/stainless steel

Click on the part number for more information about each product

Other products of Colette Industries Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Colette Industries Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

116.6hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2025 Metoree