About This Product

■Main uses

・Brightness and chromaticity distribution measurement of multiple small and medium-sized LCD panels and organic EL panels at once ・Measurement of brightness and chromaticity distribution of a single large LCD panel/organic EL panel ・Brightness distribution measurement in the lighting field ・Measurement of luminance and correlated color temperature distribution of various light emitters ・Measurement of brightness distribution of automobile meters

■Expansion of low-luminance measurable range

The minimum measurement brightness has been improved to twice that of the previous model (CA-2000), making it possible to measure 0.05cd/m2.

■Improved durability

The number of measurements that can be performed has been increased to approximately 5 times that of the previous model (CA-2000).

■High correlation with human eye sensitivity due to XYZ filter sensor

By adopting an XYZ filter with a spectral response that approximates the CIE 1931 color matching function, it has a higher correlation with the sensitivity of the human eye than RGB color separation filters used in digital video cameras. Brightness and chromaticity can be measured.

■Lens available for various targets and applications

The lineup includes three types of lenses: standard, wide-angle, and telephoto (with macro ring). You can select the optimal lens depending on the target and measurement method.

■Complete factory calibration

Extensive factory calibration is performed for each lens type, focus position, and built-in optical filter. Regardless of target size, brightness, or measurement method, it can be used for high-precision measurements immediately after purchase.

■Accurate measurement is possible even for flickering light sources such as PDP and OLED TVs.

It is possible to perform synchronous measurements on display devices and pulsed light emitters by inputting the frequency numerical value. (4~2,000Hz)

■New software CA-S25w with excellent operability is standard equipment.

Focusing assistance, positioning assistance, and automatic detection functions eliminate the need for troublesome measurement preparations.

■SDK (Software Development Kit) included as standard

We have prepared an SDK that can be used for user software development. You can efficiently create original software.

  • Product

    2D color luminance meter CA-2500

Share this product


410+ people viewing

Last viewed: 20 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

3 Models of 2D color luminance meter CA-2500

Click on the part number for more information about each product

Image Part Number Price (excluding tax) Power supply Mass Color specification mode Display mode Repeatability Brightness Repeatability Chromaticity Accuracy Brightness Accuracy Chromaticity Accuracy Measurement brightness range (using built-in ND) Number of measurement points (resolution) Measurement time Measurement size (one side of the square) Size body only Size When lens and lens hood are attached Photodetector Storage temperature and humidity range Operating temperature and humidity range Measured size at main distances (size/distance) Lens Interface Other functions
2D color luminance meter CA-2500-Part Number-CA-2500S

CA-2500S

Available upon quote

Dedicated AC adapter (100V 1.2A, 50/60Hz)

Approx. 3.5kg (with standard lens and lens hood)

XYZ, LVxy, LVu’v’, T ⊿ uv, dominant wavelength/stimulus purity

Pseudo color, chromaticity diagram, spot, cross section, color shift

0.50%

0.001

±3%

±0.005

Temperature and humidity characteristics (within the operating temperature and humidity range) Brightness: ±2% change when based on 23°C and 40% relative humidity Chromaticity: ±2% change when based on 23°C and 40% relative humidity ±0.004

0.05~100,000cd/m2

980×980 (490×490, 196×196 selectable by data management software CA-S25w)

1 accumulation: approx. 5 seconds~, 4 accumulations: approx. 6 seconds, 16 accumulations: approx. 10 seconds, 64 accumulations: approx. 25 seconds, 256 accumulations: approx. 80 seconds~

Standard lens approximately 98mm~ (variable depending on distance)

160 (width) mm x 164 (height) mm x 192 (depth) mm (height including handle: 211 mm)

223 (depth) mm

CCD image sensor (monochrome), 2/3 inch, effective pixels: 1,000 x 1,000 pixels, built-in XYZ filter (approximate to CI E1931 color matching function), built-in ND filter

0~30℃, relative humidity 70% or less/no condensation 30~35℃, relative humidity 55% or less/no condensation

10~30℃, relative humidity 70% or less/no condensation

Approx. 98mm/250mm Approx. 210mm/500mm Approx. 440mm/1,000mm Approx. 890mm/2,000mm

User-interchangeable lenses Standard, wide-angle, telephoto, low-magnification macro, high-magnification macro (low-magnification macro and high-magnification macro depend on the combination of telephoto lens and macro ring)

USB 2.0

Synchronous measurement (synchronized frequency setting possible), user calibration, integration function

2D color luminance meter CA-2500-Part Number-CA-2500W

CA-2500W

Available upon quote

Dedicated AC adapter (100V 1.2A, 50/60Hz)

Approx. 3.5kg (with standard lens and lens hood)

XYZ, LVxy, LVu’v’, T ⊿ uv, dominant wavelength/stimulus purity

Pseudo color, chromaticity diagram, spot, cross section, color shift

0.50%

0.001

±3%

±0.005

Temperature and humidity characteristics (within the operating temperature and humidity range) Brightness: ±2% change when based on 23°C and 40% relative humidity Chromaticity: ±2% change when based on 23°C and 40% relative humidity ±0.004

0.05~100,000cd/m2

980×980 (490×490, 196×196 selectable by data management software CA-S25w)

1 accumulation: approx. 5 seconds~, 4 accumulations: approx. 6 seconds, 16 accumulations: approx. 10 seconds, 64 accumulations: approx. 25 seconds, 256 accumulations: approx. 80 seconds~

Wide-angle lens approximately 145mm~ (variable depending on distance)

160 (width) mm x 164 (height) mm x 192 (depth) mm (height including handle: 211 mm)

219 (depth) mm

CCD image sensor (monochrome), 2/3 inch, effective pixels: 1,000 x 1,000 pixels, built-in XYZ filter (approximate to CI E1931 color matching function), built-in ND filter

0~30℃, relative humidity 70% or less/no condensation 30~35℃, relative humidity 55% or less/no condensation

10~30℃, relative humidity 70% or less/no condensation

Approx. 145mm/200mm Approx. 410mm/500mm Approx. 850mm/1,000mm Approx. 1,770mm/2,000mm

User-interchangeable lenses Standard, wide-angle, telephoto, low-magnification macro, high-magnification macro (low-magnification macro and high-magnification macro depend on the combination of telephoto lens and macro ring)

USB 2.0

Synchronous measurement (synchronized frequency setting possible), user calibration, integration function

2D color luminance meter CA-2500-Part Number-CA-2500T

CA-2500T

Available upon quote

Dedicated AC adapter (100V 1.2A, 50/60Hz)

Approx. 3.5kg (with standard lens and lens hood)

XYZ, LVxy, LVu’v’, T ⊿ uv, dominant wavelength/stimulus purity

Pseudo color, chromaticity diagram, spot, cross section, color shift

0.50%

0.001

±3%

±0.005

Temperature and humidity characteristics (within the operating temperature and humidity range) Brightness: ±2% change when based on 23°C and 40% relative humidity Chromaticity: ±2% change when based on 23°C and 40% relative humidity ±0.004

0.25~100,000cd/m2 When using low magnification macro 0.25~100,000cd/m2 When using high magnification macro 0.5~100,000cd/m2

980×980 (490×490, 196×196 selectable by data management software CA-S25w)

1 accumulation: approx. 5 seconds~, 4 accumulations: approx. 6 seconds, 16 accumulations: approx. 10 seconds, 64 accumulations: approx. 25 seconds, 256 accumulations: approx. 80 seconds~

Telephoto lens Approx. 115mm~ (variable depending on distance) Approx. 57mm when using low magnification macro (fixed) Approx. 27mm when using high magnification macro (fixed)

160 (width) mm x 164 (height) mm x 192 (depth) mm (height including handle: 211 mm)

224 (depth) mm When using low magnification macro 230 (depth) mm When using high magnification macro 237 (depth) mm

CCD image sensor (monochrome), 2/3 inch, effective pixels: 1,000 x 1,000 pixels, built-in XYZ filter (approximate to CI E1931 color matching function), built-in ND filter

0~30℃, relative humidity 70% or less/no condensation 30~35℃, relative humidity 55% or less/no condensation

10~30℃, relative humidity 70% or less/no condensation

Approx. 115mm/900mm Approx. 275mm/2,000mm (fixed) Approx. 420mm/3,000mm Approx. 57mm/500mm when using low magnification macro (fixed) Approx. 27mm/300mm when using high magnification macro (fixed)

User-interchangeable lenses Standard, wide-angle, telephoto, low-magnification macro, high-magnification macro (low-magnification macro and high-magnification macro depend on the combination of telephoto lens and macro ring)

USB 2.0

Synchronous measurement (synchronized frequency setting possible), user calibration, integration function

Click on the part number for more information about each product

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Luminance Color Meters Products

Other products of Konica Minolta Japan Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Konica Minolta Japan Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

6.5hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree