About This Product

■Summary

This measuring instrument has the measurement performance required in the development and inspection of high-quality displays, and is a spectral radiance meter that combines sensor technology accumulated over many years with the latest digital technology. While maintaining data compatibility with the previous model CS-2000, it has achieved accurate measurement from low to high brightness and significantly increased speed.

■Main uses

Measurement of brightness and chromaticity of various displays such as LCD and OLED

■Chromaticity accuracy guaranteed for each measurement angle

Accuracy and repeatability are guaranteed for each measurement angle in CIE1931 xy chromaticity and CIE1976 UCS chromaticity diagram u'v'.

■Short measurement time

Using a conventional spectrophotometer to perform sequence measurements involving low luminance, such as gamma measurements of displays, low luminance uniformity measurements, or I-V-L measurements of display materials, can take several hours or more, and can be difficult for both R&D and production. This is also an issue in the field of quality control. CS-2000Plus can significantly save time with the following new features.

■High speed mode

・FAST mode: Measurement integration time approximately 1/7 compared to NORMAL ・SUPER FAST1 mode: Measurement integration time approximately 1/25 compared to NORMAL ・SUPER FAST2 mode: Measurement integration time approximately 1/60 compared to NORMAL

■INTELLIGENT DARK function

By skipping dark measurements and using the correction information from the sensor to correct the fluctuations, measurement time can be shortened while maintaining high accuracy.

■Compatibility with CS-2000A/CS-2000

You can continue to use highly reliable CS-2000A/CS-2000 measurement data. Communication specifications, product shapes, and screw hole positions are also compatible, so there are no major barriers to switching to CS-2000Plus.

  • Product

    Spectroradiance meter CS-2000Plus

Share this product


190+ people viewing

Last viewed: 10 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Spectroradiance meter CS-2000Plus

Image Part Number Price (excluding tax) Power supply Brightness (Standard illuminant A) Repeatability 1° Brightness (Standard light source A) Accuracy 1° Mass Color specification mode Display wavelength interval Chromaticity (Standard illuminant A) Repeatability 1° Chromaticity (Standard illuminant A) Accuracy 1° Accuracy guaranteed brightness range (A light source) 1° Accuracy guaranteed brightness range (A light source) 0.2° Accuracy guaranteed brightness range (A light source) 0.1° Integration time Emission frequency detection Measurement angle switching Measurement angle Measurement wavelength range Measurement time Communication Measurement time Standalone Power consumption Wavelength accuracy Wavelength resolution Shortest measurement distance Minimum brightness display Minimum measurement diameter 1° Size Polarization error Operating temperature and humidity range Spectral wavelength width Interface
Spectroradiance meter CS-2000Plus-Part Number-CS-2000Plus

CS-2000Plus

Available upon quote

Dedicated AC adapter (100-240 V, 50/60 Hz)

0.4% (0.003~ cd/m2) 0.3% (0.05~ cd/m2) 0.15% (0.1~ cd/m2)

±2%

Approximately 7.0kg

Lvxy, Lvu'v', LvTΔ uv, XYZ, spectrograph, dominant wavelength, stimulus purity

1.0nm

X,y : 0.002 (0.003~0.005 cd/m2) x,y : 0.0004 (0.2~ cd/m2)

X,y : ±0.003 (0.003~0.005 cd/m2) x : ±0.0015 (0.05~ cd/m2) y : ±0.001 (0.05~ cd/m2)

0.003~5,000 cd/m2

0.075~125,000 cd/m2

0.3~500,000 cd/m2

0.005 seconds~120 seconds (NORMAL mode) 0.005 seconds~16 seconds (FAST mode)

None

Manual

1°, 0.2°, 0.1°

380~780nm

Approximately 0.08 seconds (MANUAL 33.333ms, when set to INTELLIGENT DARK)

Approximately 1~242 seconds (NORMAL mode)

Approximately 20W

±0.3 nm (Centroid wavelength Hg-Cd lamp: 435.8 nm, 546.1 nm, 643.8 nm)

0.9nm/pixel

350 mm (when using close-up lens: 55mm)

0.00002 cd/m2

Φ0.5 mm (when using close-up lens: φ0.1mm)

158 x 262 x 392 mm (body)

2% or less (400~780 nm): 1°, 3% or less (400~780 nm): 0.2°, 0.1°

5~35℃

5nm or less (half width)

RS-232C, USB 2.0

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Spectroradiometers Products

Other products of Konica Minolta Japan Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Konica Minolta Japan Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

6.6hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree