Product Image | Part Number | Price (excluding tax) | Crosshead arbitrary speed | Crosshead position accuracy | Crosshead return speed | Crosshead speed accuracy | Crosshead speed and load amount | Crosshead speed range | Crosshead table spacing | Effective stroke | Effective test width | Environmental conditions | Load method | Mass | Maximum load capacity | Noise | Power consumption/power supply/breaker capacity | Power supply | Recommended environmental conditions | Safety device Motor overload | Safety device emergency stop | Safety device overload | Safety device soft limit | Safety device stroke | Sampling rate | Size: Approximately (width x depth x height) | Stroke | Test force calibration | Test force measurement accuracy High precision type | Test force measurement accuracy standard type | Test force range |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
RTI-1310 |
Available upon quote | 0.001mm step within crosshead speed range | Within ±0.1% of indicated value. However, if the indicated value is less than 10mm, ±0.01mm | 600,1200mm/min | ±0.1% | Capable of testing up to maximum load capacity in all speed ranges | 0.001~1,200mm/min | 1,172mm | 564mm (when J-JBM-10KN is installed) | 420mm |
・Temperature: 5~40℃ ・Humidity: 20~80%RH (no condensation) |
Closed-loop digital servomechanism controlled by microcomputer | 110kg | 10kN | 65dBA or less | 475W/1.7kVA/15A | AC95~122V 1φ 50/60Hz Cable 3m end 3P plug |
・Temperature 23±2℃ ・Humidity 50%RH or less |
Attached | Attached | Attached | Attached | Attached | 0.2msec | 784×436×1,555mm | 1,100mm (1,500mm with stroke extension option) | One-touch test force calibration using the calibration circuit built into the load cell, with load cell rating identification function | Within ±0.5% of indicated value (within 1/1 to 1/500 of load cell rating) | Within ±1% of the indicated value (within the range of 1/1 to 1/500 of the load cell rating) | Full auto range (128x) | |
RTI-1225 |
Available upon quote | 0.001mm step within crosshead speed range | Within ±0.1% of indicated value. However, if the indicated value is less than 10mm, ±0.01mm | 825,1650mm/min | ±0.1% | Capable of testing up to maximum load capacity in all speed ranges | 0.001~1,500mm/min | 1,172mm | 669mm (when J-JFM-5KN is installed) | 420mm |
・Temperature: 5~40℃ ・Humidity: 20~80%RH (no condensation) |
Closed-loop digital servomechanism controlled by microcomputer | 110kg | 2.5kN | 60dBA or less | 275W/700VA/15A | AC95~122V 1φ 50/60Hz Cable 3m end 3P plug |
・Temperature 23±2℃ ・Humidity 50%RH or less |
Attached | Attached | Attached | Attached | Attached | 0.2msec | 784×436×1,555mm | 1,100mm (1,500mm with stroke extension option) | One-touch test force calibration using the calibration circuit built into the load cell, with load cell rating identification function | Within ±0.5% of indicated value (within 1/1 to 1/500 of load cell rating) | Within ±1% of the indicated value (within the range of 1/1 to 1/500 of the load cell rating) | Full auto range (128x) |