AQUA water quality measuring device cyanide ion automatic measuring device CN-180-CN-180
AQUA water quality measuring device cyanide ion automatic measuring device CN-180-Anatec Yanaco Co., Ltd.

AQUA water quality measuring device cyanide ion automatic measuring device CN-180
Anatec Yanaco Co., Ltd.


About This Product

These devices fully automate the measurement method using an ion electrode after sample pretreatment using an intermittent measurement method, and while they significantly reduce the amount of chemicals used, they further improve performance and also require automatic calibration. This is a completely redesigned product with higher grade equipment, and can be widely used as a monitor for public water areas and various industrial wastewater. ■Features - Adopts negative pressure suction method for sample water collection ・Able to meet a wide range of needs from low to high concentrations ・Can be optimized according to sample composition ■External input/output 1) Measured value voltage output DC0.000~1.000V within ±0.005V (non-insulated) 2) Measured value current output DC4.00~20.00mA±0.08mA (isolated) 3) Alarm contact output No-voltage a contact output (contact capacity 1A or less at 100V AC / 30V DC) (Changing to b contact…optional) 1.Concentration upper limit alarm 2. Sample cutting…Option 3.Cleaning water cutoff 4. Instrument abnormality 5. Power off 6. Under maintenance 4 ) External start input No-voltage a contact input (1 second or more) ■Installation conditions ・Indoor installation (installed in a building as close as possible to the water sampling point, avoiding direct sunlight) ・Temperature: 2~35℃ (30℃ or less, temperature control with air conditioner recommended) ・Humidity: 85%RH or less ・Vibration: General factory level ・Atmosphere: A place where there is no risk of corrosive gases (H2S/SO2/NH3/CL2, etc.) or sea breeze coming in directly, and where there is little dust.

  • Product

    AQUA water quality measuring device cyanide ion automatic measuring device CN-180

Share this product


20+ people viewing

Last viewed: 5 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of AQUA water quality measuring device cyanide ion automatic measuring device CN-180

Product Image Part Number Price (excluding tax) Automatic calibration mechanism Dilution/washing water conditions Display Electrode cleaning External dimensions Ground Installation space Mass Measurement method Measurement period Measurement target Measuring range Number of samples Piping diameter Power consumption Power supply Printed content Printing method Proofreading Reagent replenishment Repeatability Sample conditions Sample/reagent measurement Saved data *
AQUA water quality measuring device cyanide ion automatic measuring device CN-180-Part Number-CN-180

CN-180

Available upon quote Calibration is performed by introducing CAL-L calibration solution, and CAL-H calibration is confirmed by automatic shift.
(Standard equipped: You can choose whether to use it or not)
Tap water or purified water (for dilution measurements, use ion exchange water or pure water)
Water temperature: 2~35℃
Pressure: 0.1~0.2MPa
Usage amount: Approximately 1L/measurement
5.7 inch touch panel LCD display
Display: time, measurement status, daily report, settings, cyan recovery curve history, alarm history, etc.
Introduce cleaning water into the measurement tank and stir for cleaning. (W) 600 × (D) 600 × (H) 1,600mm (channel base type) Class D grounding (separate from power grounding) Approximately 1m of maintenance space is required at the front and approximately 0.6m at the rear. Approximately 135kg Alkaline conditions/cyanide ion electrode method Shortest measurement cycle: approximately 15 minutes
Repeat measurements, arbitrary measurements, and measurement starts using external contact signals are also possible.
Free cyanide, etc. in water 0~1 to 0~3mg/L (effective measurement range: 0.03mg/L) *Others can be manufactured 1 sample (2 or more samples available as an option) Tap water inlet: Rp1/2 (valve socket nominal diameter 13)
Drainage outlet: Rp1 (valve socket nominal diameter 25)
Sample intake port: Rc1/8 (attached union)
Approximately 100VA (at maximum load) *Excluding use of service outlet AC100V±10% 50/60Hz (built-in earth leakage breaker) Setting value, measured value, measured value bar graph, calibration value, daily report, power off,
Various alarms/individual mark printing of abnormal locations, etc. (forward/reverse order printing selectable)
Thermal printer thermal paper (paper width 60mm) (automatic winding mechanism standard equipment) Measure the standard solution in the CAL process and key in the printed value for calibration. Continuous measurement is possible for 14 days without reagent replenishment with a measurement cycle of 1 hour.
(1 month specification also available *Confirmation of installation environment etc. is required)
±3% of full scale (as equal division of logarithmic output with standard solution)
(±4% of full scale for diluted measurements)
Water temperature: 2~40℃
Pressure: 0.02~0.05MPa
Flow rate: 1~5L/min
Negative pressure suction metering method (including metering discharge method) Measurement data: Approximately 6 months
Potential history data: Approximately 1 month
Alarm history data: Latest 10 items
*This is the recordable period when the measurement interval is set to 1 hour.