Product
AQUA Water Quality Measuring Equipment Total Cyanide Automatic Measuring Device TCN-580Handling Company
Anatec Yanaco Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Control method | Dilution/washing water conditions | Display | Distillation method | External dimensions | Ground | Heating tank | Installation space | Mass | Measurement method | Measurement period | Measurement target | Measuring range | Measuring tank | Number of samples | Power consumption | Power supply | Printed content | Printing method | Proofreading | Reagent replenishment | Reagents used | Sample conditions | Sample/reagent measurement | Saved data |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
TCN-580 |
Available upon quote | Computer control |
Tap water or purified water (for dilution measurements, use ion exchange water or pure water) Water temperature: 2~35℃ Pressure: 0.1~0.2MPa Usage amount: Approximately 6L/measurement (Standard setting: Cooling water flow rate 150mL/min condition) |
5.7 inch touch panel LCD display Display: time, measurement status, daily report, settings, cyan recovery curve history, alarm history, etc. |
Hydrogen cyanide gas recovery method using negative pressure ventilation distillation under heating conditions | (W) 800 × (D) 600 × (H) 1,600mm (channel base type) | Class D grounding (separate from power grounding) | Made of quartz glass (by heater heating: with dry heating prevention sensor) | 1m required for maintenance space at the front and approximately 0.6m at the rear | Approximately 155kg | Negative pressure aeration distillation - cyanide ion electrode measurement method |
Shortest measurement cycle: approximately 30 minutes Repeat measurements, arbitrary measurements, and measurement starts using external contact signals are also possible. |
Cyanide in free and complex cyanide compounds in water |
0~1 to 5mg/L or less (effective measurement range: 0.01mg/L~) Other ranges can also be manufactured. |
Made of Pyrex amber glass (temperature-compensated concentration type using sensor) | 1 sample (2 or more samples available as an option) | Approximately 700VA (at maximum load) | AC100V±10% 50/60Hz (built-in earth leakage breaker) |
Setting value, measured value, measured value bar graph, cyan recovery curve, calibration value, daily report, power off, Printing of individual marks for various abnormalities (forward/reverse order printing selectable) |
Graphic printer thermal paper (paper width 60mm) (equipped with automatic recording paper winding mechanism as standard) |
After measuring the standard solution in the CAL process, enter the printed value for calibration using the keypad. Schedule setting is possible with the optional automatic calibration mechanism. |
Continuous measurement is possible for 14 days without reagent replenishment with a measurement cycle of 1 hour. |
Cuprous chloride/hydrochloric acid solution phosphoric acid solution sodium hydroxide solution |
Water temperature: 2~40℃ Pressure: 0.02~0.05MPa Flow rate: 1~5L/min |
Negative pressure suction metering method (including metered discharge method) |
Measurement data: Approximately 6 months Recovery curve history data: about 1 month Alarm history data: Latest 10 items |