Product
AQUA water quality measuring device TOC automatic measuring device TOC-380DHandling Company
Anatec Yanaco Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Air source conditions | Carrier gas | Control method | Detector | Display | External dimensions | Ground | Installation space | Mass | Measurement method | Measurement period | Measurement target | Measuring range | Number of samples | Piping diameter | Power consumption | Power supply | Printed content | Printing method | Proofreading | Repeatability | Sample combustion furnace | Sample conditions | Sample/reagent measurement | Saved data | Washing | Water removal | Zero correction |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
TOC-380D |
Available upon quote |
Pressure: 0.4MPa~0.9MPa (Do not contain oil or water droplets) Usage amount: Approximately 5L/min |
Instrument air (purified in equipment) | Computer control | Non-dispersive infrared analyzer |
5.7 inch touch panel LCD display Display: time, measurement status, daily report, settings, light intensity history, alarm history, etc. |
(W) 700 × (D) 600 × (H) 1,600mm (channel base type) | Class D grounding (separate from power grounding) | 1m required for maintenance space at the front and approximately 0.6m at the rear | Approximately 170kg | Combustion oxidation/non-dispersive infrared analysis method |
Shortest measurement cycle: Approx. 6 minutes (excluding cleaning time) Repeat measurements, arbitrary measurements, and measurement starts using external contact signals are also possible. |
Total organic carbon (TOC) or total carbon (TC) in water | 0~20 to 1,000mg/L *Others can be manufactured (with dilution measuring mechanism) | 1 sample (2 or more samples available as an option) |
Tap water inlet: Rp1/2 (valve socket nominal diameter 13) Drain outlet: Rp1 (valve socket nominal diameter 25) Sample inlet: Rc1/8 (attached union) |
Approximately 1KVA (at maximum load) *Excluding use of service outlet | AC100V±10% 50/60Hz (built-in earth leakage breaker) |
Setting values, measured values, measured value bar graphs, setting conditions, calibration values, daily reports, power off, Various alarms/individual mark printing of abnormal locations, etc. (forward/reverse order printing selectable) |
Thermal printer thermal paper (paper width 60mm) (automatic winding mechanism standard equipment) | Measure the standard solution in the CAL process and key in the printed value for calibration. |
±2% of full scale (with standard solution) (±3% of full scale when measuring 0 to 20 mg/L or less and adding a dilution measurement mechanism) |
Variable from 680 to 900℃ depending on the properties of the sample Made of transparent quartz glass (filled with platinum catalyst) |
Water temperature: 2~40℃ Pressure: 0.02~0.05MPa Flow rate: 1~5L/min |
Negative pressure suction metering method (including metered discharge method) |
Measurement data: Approximately 6 months End point history data: about 1 month Alarm history data: Latest 10 items |
Method of cleaning the inside of the flow path, drip pipe, catalyst, etc. using cleaning water | Electronic cooling dehumidifier | Automatic zero point method for each measurement |