Product
AQUA water quality measuring device COD automatic measuring device (reagent small quantity specification) COD-380RHandling Company
Anatec Yanaco Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Dilution/washing water conditions | Display | End point detection | External dimensions | External input/output 1 | External input/output 2 | External input/output 3 | External input/output 4 | Ground | Heating method | Installation conditions | Mass | Measurement method: Alkaline potassium permanganate method | Measurement method: Sulfuric acid acid potassium permanganate method | Measurement period | Measuring range | Number of samples | Piping diameter | Power consumption | Power supply | Printed content | Printing method | Reagent consumption: Acid method (no silver nitrate added) | Reagent consumption: Acid method (silver nitrate addition) | Reagent consumption: alkaline method | Reagent replenishment | Sample conditions | Saved data | Titration method |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
COD-380R |
Available upon quote |
Tap water or purified water (less variation in COD value) Water temperature: 2~35℃ Pressure: 0.1~0.2MPa Usage amount: 0.4~0.5L/min (for cleaning, cooling, dilution) |
5.7 inch touch panel LCD display Display: time, measurement status, daily report, settings, end point detection curve history, alarm history, etc. |
End point differential detection method of redox potential (galvanostatic polarization potentiometric method) | (W) 600 × (D) 600 × (H) 1,600mm (channel base type) | Measured value voltage output DC0.000~1.000V within ±0.005V (non-insulated) | Measured value current output DC4.00~20.00mA±0.08mA (isolated) |
Alarm contact output No-voltage a contact output (contact capacity 1A or less at 100V AC / 30V DC) (Changing to b contact…optional) Concentration upper limit alarm Sample cutting...option Washing water cutoff Instrument abnormality End point detection Power off Under maintenance |
External start input No-voltage A contact input (1 second or more) | Class D grounding (separate from power grounding) | Boiling water circulation method (oil circulation method available as an option) |
Indoor installation (installed in a building as close as possible to the water sampling point, avoiding direct sunlight) Temperature: 2~40℃ Humidity: 85%RH or less Vibration: General factory level Atmosphere: A place free of corrosive gases (H2S/SO2/NH3/CL2, etc.) |
Approximately 120kg | Alkaline method: When the chloride ion concentration of the sample (+ dilution water) is approximately 2,000 mg/L or more |
Acidic method (no silver nitrate added): When the chlorine ion concentration is low and the catalytic effect of silver salt is less affected Acidic method (silver nitrate addition): When the chloride ion concentration of the sample (+ dilution water) is approximately 2000 mg/L or less |
Schedule setting method (repeat measurement/arbitrary measurement) Measurement can also be started using an external start signal | Specify one range from 0~20,30,40,50,100,200,300,400mg/L (0~500mg/L or more is possible as an option) | 1 sample (2 or more samples available as an option) |
Tap water inlet: Rp1/2 (valve socket nominal diameter 13) Drain outlet: Rp1 (valve socket nominal diameter 25) Sample inlet: Rc1/8 (attached union) |
Approximately 700VA (at maximum load) | AC100V 50/60Hz (built-in earth leakage breaker) | Setting values, measured values, manual analysis conversion values (according to y=a+bx settings), measured value bar graphs, calibration values, daily reports, power off, various alarms/abnormal location individual marks printing, etc. (forward/reverse order printing selectable) ) | Thermal printer thermal paper (paper width 60mm) (automatic winding mechanism standard equipment) |
Potassium permanganate solution 7mL Sodium oxalate solution 5mL Sulfuric acid solution 5mL |
Potassium permanganate solution 7mL Sodium oxalate solution 5mL Sulfuric acid solution 5mL Silver sulfate solution 2.5mL Ammonia water approx. 120mL |
Potassium permanganate solution 7mL Sodium oxalate solution 5mL Sulfuric acid solution 5mL Sodium hydroxide solution 5mL |
Continuous measurement is possible for 14 days without reagent replenishment with a measurement cycle of 1 hour. |
Water temperature: 2~40℃ Pressure: 0.02~0.05MPa Flow rate: 1~5L/min |
Measurement data: Approximately 6 months End point history data: about 1 month Alarm history data: Latest 10 items |
Slow injection continuous titration method using burette |