Product
AQUA water quality measuring device COD automatic measuring device COD-3100Handling Company
Anatec Yanaco Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Dilution/washing water conditions | Display | End point detection | External dimensions | External input/output 1 | External input/output 2 | External input/output 3 | External input/output 4 | External input/output 5 | Ground | Heating method | Installation conditions | Mass | Measurement method: Alkaline potassium permanganate method | Measurement method: Sulfuric acid acid potassium permanganate method | Measurement period | Measuring range | Number of samples | Piping diameter | Power consumption | Power supply | Printed content | Printing method | Reagent consumption: Acid method (no silver nitrate added) | Reagent consumption: Acid method (silver nitrate addition) | Reagent consumption: alkaline method | Reagent replenishment | Sample conditions | Saved data | Titration method |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
COD-3100 |
Available upon quote |
Tap water or purified water (less variation in COD value) Water temperature: 2~35℃ Pressure: 0.1~0.4MPa Usage amount: 1.5L/measurement (for cleaning and dilution) |
7 inch lcd display | End point differential detection method of redox potential (galvanostatic polarization potentiometric method) | (W) 600 × (D) 600 × (H) 1,600mm (channel base type) | Measured value voltage output DC0.000~1.000V within ±0.005V (non-insulated) |
Measured value current output DC4~20mA±0.08mA (insulated) Load resistance 500Ω or less *Equipped with 1 output as standard Measured value output is a hold signal (until next measured value output) |
Alarm contact output No-voltage A contact output (1A or less when contact capacity is 30V DC) Concentration upper limit alarm Sample cutting *optional Washing water cutoff Instrument abnormality End point detection Power off |
Maintenance contact output (MAINT.) No-voltage a contact output (contact capacity 1A or less when DC30V) | External start input No-voltage A contact input (1 second or more) |
Class D grounding (separate from power grounding) (No sudden voltage fluctuations) |
Boiling water circulation method (oil circulation method available as an option) |
Indoor installation (installed in a building as close as possible to the water sampling point, avoiding direct sunlight) Temperature: 2~40℃ Humidity: 85%RH or less Vibration: General factory level Atmosphere: A place free of corrosive gases (H2S/SO2/NH3/CL2, etc.) |
Approximately 145kg | Alkaline method: When the chloride ion concentration of the sample (+ dilution water) is approximately 2,000 mg/L or more |
Acidic method (no silver nitrate added): When the chloride ion concentration is low and the catalytic effect of silver salt is less affected Acidic method (silver nitrate addition): When the chloride ion concentration of the sample (+ dilution water) is approximately 2000 mg/L or less |
Shortest measurement cycle: approximately 45 minutes Repeat measurements, arbitrary measurements, and measurement starts using external contact signals are also possible. |
Specify one range from 0~20,30,40,50,100,200,300,400mg/L (0~500mg/L or more is possible as an option) | 1 sample (2 or more samples available as an option) |
Tap water inlet: Rp1/2 (valve socket nominal diameter 13A) Drain outlet: Rp1 (valve socket nominal diameter 25A) Sample inlet: Rc1/8 (attached union) |
Approximately 700VA (at maximum load) | AC100V 50/60Hz (built-in earth leakage breaker) | Setting values, measured values, manual analysis conversion values (according to y=a+bx settings), measured value bar graphs, calibration values, daily reports, power off, various alarms/abnormal location individual marks printing, etc. (forward/reverse order printing selectable) ) | Graphic printer thermal paper (paper width 58mm) (automatic winding mechanism included as standard) |
Potassium permanganate solution 13.5mL Sodium oxalate solution 10mL Sulfuric acid solution 10mL |
Potassium permanganate solution 13.5mL Sodium oxalate solution 10mL Sulfuric acid solution 10mL Silver sulfate solution 5mL Ammonia water approx. 170mL |
Potassium permanganate solution 13.5mL Sodium oxalate solution 10mL Sulfuric acid solution 10mL Sodium hydroxide solution 10mL |
Continuous measurement is possible for 14 days without reagent replenishment with a measurement cycle of 1 hour. |
Water temperature: 2~40℃ Pressure: 0.02~0.05MPa Flow rate: 1~5L/min |
Measurement data: Approximately 6 months End point history data: about 1 month Alarm history data: Latest 10 items |
Slow injection continuous titration method using burette |