Product
Computer-controlled twin capillary rheometer R6000Handling Company
ITS Japan Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Capillary die lineup | Capillary die precision | Measurement Force measurement | Measurement Measurement accuracy | Measurement piston control range | Measurement pressure sensor | Measurement selection range | Measurement sensor connection channel | Measurement temperature sensor | Mechanism External dimensions/weight | Mechanism Required power supply | Mechanism frame strength | Mechanism measuring barrel | Mechanism safety mechanism | Mechanism structure | RheoSolve Software Specification Recommended System | RheoSolve Software Specifications Control Functions | RheoSolve software specifications Display/output functions | Temperature control Temperature control range | Temperature control Temperature measurement accuracy | Temperature control material | Temperature control temperature sensor |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
R6000 |
Available upon quote |
Length: 0.25mm~40mm Diameter: 0.25mm~5mm Incident angle: 60°~180° |
Diameter: within ±5μm Length: Within ±10μm |
Transducer in crosshead (standard 50kN) | ±0.25% of measurement range |
0.01mm/min~500mm/min Drive motor encoder 10,000 (resolution: 0.02μm) Error less than 50μm for piston full travel. |
Direct detection of pressure loss within each bore. | 3.5・5・7・10・20・35・50・70・100・140・200Mpa | 5 channels, 4-20mA 16bit data conversion | Platinum sensor (according to DIN43760) |
730mm (width) x 800mm (depth) x 2,400mm (height) Weight: 400kg |
Single phase 220VAC 50/60Hz 16A | 290kN |
Twin bore type ◎Each φ15mm standard (optional φ12mm, φ20mm) Bore length: 280mm (with option to tilt the barrel forward by 28° for cleaning) |
Protective screen door (interlocking) Crosshead force transducer, motor overcurrent, emergency stop switch, exhaust mechanism compatible, machine safety standards 89/392/EEC and 91/368/EEC certified |
Sturdy H frame structure with two φ65mm bars. |
Pentium3 1GHz or higher 512MB memory 40GB hard disk 10Base T network port for connection to rheometer. |
Crosshead control, temperature control (3 zone division), measurement condition setting (temperature equilibrium and pressure value) | Graph display of shear viscosity, extensional viscosity, etc., real-time graph display of sensor detection values, list display of measurement conditions, list display of numerical data, data output in Microsoft Excel, CSV, ASCII format. | Room temperature +10℃~400℃ | Less than ±0.5℃. | 3-zone temperature measurement and control. | Platinum sensor (according to DIN43760) |
High load type Iprexor series 2000N
ITS Japan Co., Ltd.
10+ people viewing
Last viewed: 14 hours ago