Product
Medical AC DC Open Frame Quad Output FPM110Handling Company
FranMar International Inc.Categories
Product Image | Part Number | Price (excluding tax) | Cooling | Derating | EN55011 | EN61000-3-2 | EN61000-3-3 | EN61000-4-2 | EN61000-4-3 | Efficiency | Hold up time | Input current | Input frequency | Input voltage | Inrush current | Leakage current | Line regulation | MTBF | Maximum output power | Operating temperature | Output voltage/current | Overcurrent protection | Overvoltage protection | PFD | Relative humidity | Ripple and noise | Storage temperature | Switching frequency | Temperature coefficient | Touch current | Transient response | Withstand voltage |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
FPM110-S033 |
Available upon quote | 72-110 watts continuous output with 25CFM forced air cooling or 60-80 watts with convection cooling | Linear derating from 100% at +50°C to 50% at +70°C | Class B was conducted and Class B was radiated | Harmonic distortion, class A | Flickering lines | ESD, ±15KV air and ±8KV contacts | Radiation resistant, 10V/m | Min. 70% for single output models with Vo≧12V, min. 65% for other models | 12ms minimum at 110VAC | 3.20A (rms) at 115VAC; 1.80A (rms) at 230VAC | 47~63Hz | 85~264VAC | 15A@115VAC or 30A@230VAC, cold start at 25℃ | Up to 220μA@264VAC, 63Hz | Maximum ±0.5% at full load |
Calculated according to MIL-HDBK-217F, 400,000 hours at full load at 25°C ambient temperature EMC performance (EN60601-1-2) |
See rating chart. | -10℃~+70℃ | See rating chart. | All outputs protected against short circuit conditions | Provided only on output #1. Set to 112~132% of nominal output voltage | TTL logic is high for normal operation and low when input power is lost. This signal will appear for at least 1 ms before the V1 output drops 5% below its nominal value. | 5% to 95% non-condensing | 1% peak-to-peak maximum | -40℃~+85℃ | 20~250KHz, depends on load and line | Total output ±0.04%/℃ maximum | Up to 110μA@264VAC, 63Hz | Excursion up to 4% or more for all models, recovery to 1% of final value within 500us after 25% step load change | 5,600VDC from input to output (2MOPP); 2,100VDC from input to ground (1MOPP); 700VDC from output to ground |