Product
Portable relative humidity/temperature calibrator HygroGen2-SHandling Company
PST Japan Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Application example | Autonomous driving program | C.E. | Chamber effective volume | Control | Desiccant | Occurrence range | Order code | Power supply | Probe specifications | Response speed | Sample air extraction | Size | Stability at equilibrium | Temperature distribution | USB port | Uncertainty | Water level display | Water supply for humidification | Water tank | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
HygroGen2-S |
Available upon quote | HG2-S |
Enter temperature/humidity set points and operating time, and set up to 200 points per program. Up to 20 programs can be registered |
Safty:EN61010-1:2001 EMC: EN61326-1:2006, EN61000-6-1:2007 |
Approximately 1.5L |
Humidity: split flow method Dry air: dehumidification using desiccant Temperature: Peltier element |
Molecular sieve |
Humidity: 5~95%RH Temperature: 0~+60℃ |
HG2-S | 110~240VAC, 50~60Hz, 3A |
Humidity: ±0.8%rh (23 ±5℃) ±2%rh (0~60℃) Temperature: ±0.1℃ (23 ±5℃) ±0.3℃ (0~60℃) |
Humidity: 35%rh~85%rh approximately 5 minutes T90 Typical Temperature: 23℃~30℃ about 5 minutes T90 Typical |
IN/OUT Φ6mm | 450x406x205mm (Max.) |
Humidity: <0.1%rh Temperature: <0.01℃ |
<0.05℃ (15~50℃) <0.1℃ (5~60℃) ±0.15℃ (0℃) |
7 ports on the front, 2 ports on the back |
Humidity: ±1.5%rh @ 23℃ Temperature: ±0.15℃, 15~50℃ |
Alarm function, water level displayed on monitor | Uv sterilized water | Approximately 80ml (use distilled or pure water) | Approximately 13 kg |