Zirconia oxygen concentration meter XZR200Handling Company
PST Japan Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | Electrical output/input power supply | Electrical output/input power consumption | Electrical output/input Output range | Electrical output/input Digital communication | Electrical output/input Analog output | Weight | Housing material | Welding part material | Size | Ambient temperature | Protection requirements | Probe length | Process connection | Sensor Measuring gas | Sensor accuracy (0-25%) | Sensor accuracy (0-100%) | Sensor measurement range | Sensor response speed (T90) | Sensor output resolution | Sensor repeatability | Sensor background gas | Sensor Sensor type | Sensor sample temperature | Sensor sample flow rate | Sensor sample pressure | Sensor sample flow effect | Warm up time |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
XZR200 |
Available upon quote |
24VDC, ±10% |
Max2.0A @24VDC |
0-25%, 0-100% |
RS232 (cannot be used when 4-20mA is selected) |
4-20mA, 1-10VDC |
<0.5kg |
Waterproof die-cast aluminum housing |
Stainless steel |
H52 x w75 x d80mm *excluding probe |
−10-+85℃ |
IP65 |
210mm or 400mm Φ12mm |
12mm Swagelok® compression fitting |
Oxygen |
<0.5% O2 |
<1% O2 |
0-25% or 0-100% |
<5 seconds |
0.01V, 0.01mA, 0.01% O2 |
<0.5% |
Air, N2, CO2, Ar, combustion gas |
Michelle zirconia type |
~400℃ |
0-10m/sec |
Atmosphere |
±0.1% O2 (0-1L/min) |
Approximately 10 minutes |