Moisture analyzer in natural gas OptiPEAK TDL600Handling Company
PST Japan Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | Power requirement | Electrical connection | Weight | Accuracy | Housing | Environmental conditions | Measuring range | Measurement technology | Unit of measurement | Detection limit | Calibration Recommended calibration interval | Calibration Factory default | Response speed | Outlet pressure | Reproducibility | Input pressure | Supply voltage | Data logging | Digital communication | Display | Sample flow rate | Sampling system material | Gas connection | Analog signal |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
TDL600 |
Available upon quote |
Indoor: 80W |
3 x M20 cable gland connections |
40kg (excluding sampling system) |
±1% of reading >100 ppmV |
Aluminum alloy, explosion-proof, polyester |
Indoor version: +10 ℃ ~ +45 ℃ |
1 ppmV ~ 1,000 ppmV |
TDLAS |
PpmV, lb/MMSCF, mg/Nm3, |
1 ppmV |
Depends on measurement environment requirements or system quality |
3 points, NPL and NIS traceable |
Optical response/0.2 seconds ℃ Update display interval/2~3 seconds |
Sample cell vent 0.7~1.4 bara |
<1 ppmV |
Up to 1,450 psig (100 barg) |
110VAC or 230VAC 50/60Hz |
Logs all process variables for selectable sample periods (10 seconds to 1 day) |
AC, 3A (no voltage contact) RS485 ModBus RTU |
4.3” color LCD, touch screen operation |
1 NL/min (sample cell) |
304L or 316L stainless steel |
Polyester coating, IP66, |
Input: 2 x 4–20 mA user selectable |